Features
& Benefits
THS730A/THS720A/THS720P
- 200 and 100 MHz
Bandwidth
- 1 GS/s and 500 MS/s Sampling Rates per Channel
- Two
Channels and Dual Digitizers
- External Trigger
- 3-3/4 Digit
DMM with Datalogger
- Roll Mode
- Extra Bright Backlit Display
- 8 ns
Glitch Capture
- Independently Floating Isolated Scope and DMM Channels
for Safety
- Full Floating Measurement Capability up to 600 VRMS
- Advanced
Trigger - Delay, Pulse, and Video (Line Count and Field Select)
- 21
Automatic Measurements
- RS-232 Programmability/Communications
- NiCd
Battery and AC Adapter
- Deluxe Soft Carrying Case
THS720P
Only
- Measures Harmonics up to 31st (Fundamental from 30 Hz to
450 Hz)
- Automatic Power Measurement with Statistics
- Advanced
Trigger - Delay, Pulse, Video (Line Count and Field Select), PWM Motor Drive
- 1
kV High-voltage Probes
Applications
THS730A/THS720A/THS720P
- Field
or Bench Electronic Troubleshooting
- Floating (Nongrounded) System
Measurements
- Industrial Electronic Equipment Troubleshooting (THS720P)
- Power
Quality Surveys (THS720P)
THS700 Series
Handheld
Battery Operated Oscilloscopes
The TekScope® THS700 Series Oscilloscope/DMM
combines a full-featured real-time oscilloscope with a True RMS digital multimeter
in rugged, battery-operated instruments. Scope and meter modes can operate
simultaneously and independently on the same or separate signals. The high-resolution,
backlit display and pop-up menus make it easy for users to take full advantage
of the instrument's many features. These include cursors, video trigger, voltage
and resistance measurements, and storage of waveforms, data, and instrument
setups. In addition, they offer the most comprehensive triggering capabilities
- external trigger and extensive triggering including delay, pulse width,
and video.
The THS700A Series is ideal for electronic applications
- at the bench or at the site.
The THS720P Oscilloscope/DMM includes
features to address the needs of electrical/power electronics measurements.
THS720P.
The
Only Safe Measurement Solution for Power Electronics
The THS720P TekScope
IsolatedChannel™ Oscilloscope/DMM is ideal for electrical/power electronics
applications.
The THS720P includes features specifically for electrical/power
electronics measurements which allow testing and verifying correct operation
of motors, checking transformer efficiency, verifying power-supply performance,
and measuring the effect of neutral current. These features include harmonic
analysis, power measurements and statistics and PWM motor drive triggering.
It also contains the powerful features of a modern oscilloscope that enable
troubleshooting and verification of complicated electronic control circuits
controlling the high-voltage power-electronics circuitry.
The THS720P
shares measurement features with the THS730A, THS720A Oscilloscope/DMMs, which
are ideal for electronic applications.
Characteristics
Oscilloscope
Functions
|
|
THS730A
|
THS720A
|
THS720P
|
|
Bandwidth
|
200 MHz
|
100 MHz
|
100 MHz
|
|
Sample Rate (Each Channel)
|
1 GS/s
|
500 MS/s
|
500 MS/s
|
|
Time/Division Range
|
2 ns to 50 s/div
|
5 ns to 50s/div
|
5 ns to 50s/div
|
|
Power Measurements
|
N
|
N
|
Y
|
Oscilloscope Functions
Channels - Two.
Sensitivity
- 5 mV to 50 V/div (to 500 V/div with 10X probe).
Position
Range - ±10 div.
DC Gain Accuracy - ±2%.
Vertical
Resolution - 8-Bits.
Record Length - 2500 points.
Horizontal
Accuracy - ±200 ppm.
Roll Mode - ≥ 0.5 s/div.
Autorange
- User-selectable.
Trigger Modes - Auto, Normal.
Trigger
Types - Edge, Pulse, Video, External.
Video Trigger Formats
and Field Rates - Odd field, even field, and line.
Motor Trigger
(THS720P only) - Triggers on 3- and 5-level pulse-width modulated power
signal.
Harmonics (THS720P only) - Up to 31st (30 Hz-450 Hz).
External
Trigger Input - 5 MHz TTL compatible.
Waveform Processing - Add,
Subtract, Multiply, Calculate Watts = V x I.
Waveform Storage - 10
waveforms.
Acquisition Modes - Sample, Envelope, Average, Peak
Detect.
Cursor Measurements - delta Volts, delta Time, 1/deltaTime
(Hz), Degree (phase).
Cursor Types - Horizontal Bars, Vertical
Bars, Paired (volts @ time).
Display System -
Interpolation:
Sin(x)/x.
Format: YT and XY.
Mode: Vector, Dot, Vector Accumulate,
Dot Accumulate.
Automatic Measurements
|
Period
|
Frequency
|
|
+Width
|
-Width
|
|
Rise Time
|
Fall Time
|
|
+Duty Cycle
|
-Duty Cycle
|
|
+Overshoot
|
-Overshoot
|
|
High
|
Max.
|
|
Low
|
Min.
|
|
Peak-to-Peak
|
Amplitude
|
|
Mean
|
RMS
|
|
Cycle Mean
|
Cycle RMS
|
|
Burst Width
|
|
Power Measurements (THS720P Only)
|
W
|
True power
|
|
VA
|
Apparent power
|
|
VAR
|
Reactive power
|
|
V
|
Volts (RMS, Peak)
|
|
A
|
Amps (RMS, Peak)
|
|
THD-F
|
Total harmonic distortion as a percentage of the fundamental
|
|
THD-R
|
Total harmonic distortion of the RMS of the input signal
|
|
PF
|
Power factor
|
|
DPF
|
Displacement power factor
|
|
PHI
|
Phase difference between the voltage and current
|
DMM Specifications
DC Voltage Ranges - 400.0
mV to 880 V.
DC Volts Accuracy - ±(0.5% of reading + 5 counts).
True
RMS AC Voltage Ranges - 400.0 mV to 640 V.
Maximum Float Voltage
- 600 VRMS CAT II, each channel (probe dependent).
Resolution
- 4000 count, 3-3/4 digits.
AC Volts Accuracy - ±(2% of
reading + 5 counts).
Resistance Ranges - 400.0 Ω to 40.00 MΩ.
Resistance
Accuracy - ±(0.5% of reading + 2 counts); 40 MΩ: ±(2% of reading + 5 counts).
Diode
Test Range - 0 to 2 V.
Continuity Check - Audible tone
when < 50 Ω.
Modes - Min, Max, Delta, Max-Min, Avg, Hold.
Nonvolatile
Storage - 10 DMM screenshots.
External Trigger Input - 5
MHz TTL compatible.
Vertical Zoom Capability - 2X, 5X, 10X.
dB
Scale - Selectable, referenced from 1 mV to 10 V.
dBm Scale
- Selectable, referenced from 50 Ω to 600 Ω.
General Specifications
Setups
- 10 front-panel setups.
Safety - UL3111-1, CSA1010.1,
EN61010-1, IEC61010.1.
EMC - Meets Directive 89/336/EEC.
Power
- NiCd rechargeable battery pack with AC adapter (both included).
Battery
Life - Approx. 2 hours from full charge.
Display - Backlit
LCD.
Display Resolution - 320 x 240.
Physical Characteristics
|
Dimensions
|
mm
|
|
Weight
|
kg
|
|
Width
|
177
|
|
Height
|
217
|
|
Depth
|
51
|
|
Net
|
1.45
|
Ordering
Information
THS730A, THS720A
Handheld Battery Operated Oscilloscope/DMM.
Includes: Two P6117 10X
Passive Probes, User Manual, Quick Reference Manual, Standard Meter Lead Set,
NiCd Rechargeable Battery Pack, AC Power Adapter, Soft Carrying Case, Cable
and Adapters for RS-232, NIST-traceable Certificate of Calibration.
Please
specify power plug and manual version when ordering.
THS720P
Handheld
Battery Operated Oscilloscope/DMM.
Includes: Two P5102 10X
Passive Probes, User Manual, Quick Reference Manual, Standard Meter Lead Set,
NiCd Rechargeable Battery Pack, AC Power Adapter, Soft Carrying Case, Cable
and Adapters for RS-232, NIST-traceable Certificate of Calibration.
Language
Options
Opt. L0 - English User Manual
Opt. L1 - French
User Manual
Opt. L2 - Italian User Manual
Opt. L3
- German User Manual
Opt. L4 - Spanish User Manual
Opt.
L5 - Japanese User Manual
Opt. L6 - Portuguese User Manual
Opt.
L7 - Simplified Chinese User Manual
Opt. L8 - Traditional
Chinese User Manual
Opt. L9 - Korean User Manual
Opt.
L10 - Russian User Manual
Power Plug Options
Opt.
A0 - US Plug, 115 V, 60 Hz
Opt. A1 - Euro Plug, 220V, 50
Hz
Opt. A2 - UK Plug, 240V, 50 Hz
Opt. A3 - Australian
Plug, 240V, 50 Hz
Opt. A6 - Japanese Plug, 100V, 110/120 Volt,
60 Hz
Recommended Accessories
Also see Accessories.
P5102 - 10X 100 MHz Passive Probe for higher voltage
application: 1000VRMS Tip-to-reference; 600 VRMS Reference-to-earth
Ground (float).
A621, A622 - Current
Probes.
P6117 (replacement) - 10X Passive
Probe.
Battery Charger -
THS7CHG.
Rechargeable
Battery (replacement) -
THS7BAT.
Standard
Meter Lead Set (replacement) - 012-1482-00.
Programming Manual
- 070-9751-01.
Service Manual - 070-9752-01.
Soft
Carrying Case (replacement) - 016-1399-01.
Software
WSTRO - WaveStar
software for Oscilloscopes, Windows 95/98/NT application for waveform capture,
analysis, documentation and control form your PC.
Warranty
Three
years (excluding probes).
|
|
Product(s) are manufactured in ISO registered facilities.
|
3MW-10566-6, 14-JUN-2006
|