Features & Benefits
- Bandwidths up to 100 MHz
- Up to 5,600 V Differential (DC + pk AC)
- Up to 2,200 V
Common (RMS)
- Overrange Indicator
- Safety Certified
- Switchable Attenuation
- Switchable Bandwidth Limit
Applications
- Floating Measurements
- Switching Power Supply Design
- Motor Drive Design
- Electronic Ballast Design
- CRT Display Design
- Power Converter Design and Service
- Power Device Evaluation
P5200
The P5200 can be used with
any oscilloscope and enables users to safely make measurements
of floating circuits with their oscilloscope grounded. The P5200 Active
Differential Probe converts floating signals to low-voltage ground-referenced
signals that can be displayed safely and easily on any ground-referenced
oscilloscope.
WARNING: For safe operation, do not use
the P5200 High-voltage Differential Probe with oscilloscopes that
have floating inputs (isolated inputs), such as the Tektronix TPS2000
Series oscilloscopes and THS700 Series oscilloscopes. The P5200 High-voltage
Differential Probe requires an oscilloscope or other measurement instrument
with grounded inputs.
P5210
The P5210 is a Differential Probe that is capable of measuring
floating voltages up to 5,600 V (DC + pk AC) safely and has
a bandwidth up to 50 MHz. It is supplied with two sizes of hook tips
and has an overrange visual and audible indicator which warns the
user when they are exceeding the linear range of the probe. It can
be used with Tektronix TEKPROBE™ interface oscilloscopes directly
or with any oscilloscope with the use of the 1103 TEKPROBE® Power Supply.
The P5205 is a 100 MHz Active Differential
Probe capable of measuring fast rise times of signals in floating
circuits. This 1,300 V differential probe can safely measure
voltages in IGBT circuits such as motor drives or power converters.
It is specifically designed to operate on Tektronix oscilloscopes
with TEKPROBE® interface.
P5200
P5205
P5210
Characteristics
|
Characteristic
|
P5200
|
P5205
|
P5210
|
|
Bandwidth (-3 dB)
|
25 MHz
|
100 MHz
|
50 MHz
|
|
Differential Mode Voltage (RMS or DC)
(between positive
and negative input leads)
|
1000 V CAT II
|
1000 V CAT II
|
4400 V
|
|
Differential Mode Voltage (Peak)*1
(between
positive and negative input leads)
|
1300 V
|
1300 V
|
5600 V
|
|
Common Mode Voltage
(RMS or DC + pk AC)
(between
positive lead and ground or negative lead and ground)
|
1000 V CAT II
600 V CAT III
|
1000 V CAT II
600 V CAT III
|
2200 V CAT II
1000 V CAT III
|
|
Attenuation (switchable)
|
50x/500x
|
50x/500x
|
100x/1000x
|
|
CMRR at 60 Hz / 1 MHz (typical)
|
80 dB / 50 dB
|
80 dB / 50 dB
|
80 dB / 50 dB
|
|
Input R (each input)
|
4 MΩ
|
4 MΩ
|
8 MΩ
|
|
Input C (each input)
|
7 pF
|
7 pF
|
7 pF
|
|
Power Source
|
AC wall adapter
|
TEKPROBE Power Supply
|
TEKPROBE Power Supply
|
|
Safety (all)
|
UL3111-1/CSA1010.1/CSA1010.2.031/IEC61010-2-031/EN61010-2-031
|
|
Compatible Products
|
ALL*3
|
TDS400/500/600/700/ TDS3000B/5000B/6000B*2/7000B*2/ CSA7000B*2 and 11000 Series
|
TDS400/500/600/700/ TDS3000B/5000B/6000B*2/7000B*2/ CSA7000B*2 and 11000 Series
|
*1 The RMS value of the waveform shall
not exceed the Common Mode Voltage (RMS) for either lead.
*2 TDS7154B, TDS7254B, TDS7404B, TDS6404, TDS6604, CSA7154,
and CSA7404B require a TCA-1MEG Buffer Amplifier.
*3
WARNING: For safe operation, do not use the P5200 High-voltage
Differential Probe with oscilloscopes that have floating inputs (isolated
inputs), such as the Tektronix TPS2000 series oscilloscopes and THS700
series oscilloscopes. The P5200 High-voltage Differential Probe requires
an oscilloscope or other measurement instrument with grounded inputs.
Ordering Information
P5200
Active Differential
Probe.
Includes: 2 crocodile clips; 2 plunger clips;
9 VDC power supply; manual (070-9018-01). Please specify power cord
when ordering.
Power Plug Options (P5200 only)
Opt. A0 - US Plug, 115 V, 60 Hz.
Opt. A1 - Euro Plug, 220 V, 50 Hz.
Opt. A2 - UK Plug, 240 V,
50 Hz.
Opt. A3 - Australian Plug, 240 V, 50 Hz.
Opt. A5 - Switzerland Plug, 220 V, 50 Hz.
Opt. A6 - Japanese Plug, 100 V, 110/120 Volt, 60 Hz.
See Power Sources page for additional description.
P5205
Active Differential Probe.
Includes: 2 each plunger clamps; 2 each crocodile clips; 2 each extender leads;
2 each lead adapters; 2 each plunger hook clips; soft case; manual
(070-9472-01).
P5210
Active Differential Probe.
Includes: 2 each large hook clips; 2 each small hook
clips; 2 each probe heads; soft case; manual (070-9841-01).
Service
Opt. C3 - Calibration Service 3 Years.
Opt. C5 - Calibration Service 5 Years.
Opt.
D1 - Calibration Data Report.
Opt. D3 - Calibration
Data Report 3 Years (with Opt. C3).
Opt. D5 - Calibration
Data Report 5 Years (with Opt. C5).
Opt. R3 - Repair
Service 3 Years.
Opt. R5 - Repair Service 5 Years.
Please see the Customer Support page.
Accessory Pack (P5200, P5205)
Order 020-2106-xx.
Includes: 2 each plunger clamps; 2 each crocodile clips.
Accessory pack (P5200, P5205)
Order 020-2140-xx.
Includes: 2 each extender leads; 2 each lead adapters;
2 each plunger hook clips.
Accessory Pack (P5210)
Order 020-2195-xx.
Includes: 2 each large hook
clips; 2 each small hook clips; 2 each probe heads.
Soft
case
Order 016-1461-xx.
Power Supply (P5205,
P5210 only)
TEKPROBE® Power Supply. Order 1103.