Features & Benefits
- Bandwidths of 15 GHz (TDS6154C),
12 GHz (TDS6124C), 8 GHz (TDS6804B), and 6 GHz (TDS6604B)
- Rise Times to 19 ps 20-80% (28 ps 10-90%) on TDS6154C Typical Rise
Time, with Channel-matched, User-selectable DSP
- 40 GS/s Real-time
Sample Rate on Two Channels*1, 20 GS/s Real-time Sample
Rate on All Four Channels with 500 fs/sample Interpolated Points
- Up to 64 Megasamples on Two Channels*1, up to 32 Megasamples
Record Length on All Four Channels with MultiView Zoom™ Function for
Quick Navigation
- MyScope® Custom-control Windows
Enhance Productivity
- Right-mouse-click Menus for Exceptional
Efficiency
- Pinpoint™ Triggering Provides the Most Flexible
and Highest Performance Triggering, with Over 1400 Combinations to
Address Virtually Any Triggering Situation
- Serial Pattern
Triggering up to 3.125 Gb/s with 8b/10b Protocol Triggering
- Serial Data Analysis and Compliance at Rates of 6.25 Gb/s and Above
- OpenChoice® Software with Microsoft Windows XP OS
Delivers Built-in Networking and Analysis
- Technology-specific
Software Solutions Provide Built-in Domain Expertise for Serial Data,
Jitter, Ethernet, DVI, USB 2.0, Communications, and Power Measurements
- System Includes: Dual-processor System (2.8 GHz Pentium 4 and
583 MHz PowerPC), High-resolution XGA Display, Front-panel CD-R/W,
Front-panel USB 2.0 Port, and 1000BaseT Network Connection
Applications
- Signal Integrity, Jitter,
and Timing Analysis
- Validation, Debug, Characterization,
and Compliance of Next-generation Digital Designs
- Computer,
Datacom, Storage-area Network Equipment Designs, and High-speed Backplanes
- High-energy Physics Measurements and Data Acquisition
*1 For C model versions only.
Uncompromised
Performance Oscilloscope Solution – Probing, Acquisition, Analysis,
Compliance, and Debug
TDS6000B/C Series digital storage oscilloscopes
provide unprecedented performance along with a complete feature set
designed to address design validation, debug, and compliance challenges
of next-generation computer, datacom, and communications equipment.
High bandwidth, high sample rate, and deep memory also provide the
ideal solution for data acquisition applications.
Uncompromised
Acquisition
You won’t need to need to trade off bandwidth,
record length, and sample rate for your serial data measurement and
analysis needs. The TDS6000C DSOs provides acquisition architecture
with 40 GS/s maximum sample rate and 64 Megasamples record length
on two channels (20 GS/s and 32 Megasamples on each of the four channels
simultaneously), for the acquisition power you need. They provide
the ultimate combination of bandwidth, sample rate, and record length
for the fastest signals. The TDS6154C provides matched 15 GHz performance
across any two channels using advanced, Tektronix-proprietary DSP
enhancement, important for high-speed channel-channel measurements.
The user-selectable DSP filter on each channel provides magnitude
and phase correction, plus extension of the analog bandwidth to 15 GHz
for more accurate signal fidelity on high-speed measurements – easily
capturing the fifth harmonic of 3.0 GHz embedded clocks used in next-generation
6.0 Gb/s serial data standards, and even the third harmonic of 5 GHz
clocks being developed for future systems. The DSP filter on each
channel can also be switched off to take advantage of true 12 GHz
analog bandwidth for applications needing the highest available raw
data capture.
MyScope® Custom-control Windows
MyScope control windows allow you to build your own control windows
with only the controls, features, and capabilities that you care about
and are important in your job. For the first time you can create your
own personalized “toolbox” of oscilloscope features. No longer do
you need to search through menus for features or relearn how to drive
the oscilloscope after a break from the lab. MyScope control windows
are easily created in a matter of minutes using a simple, visual,
drag-and-drop process. Once created, these customized windows are
easily accessed through a dedicated MyScope button and menu selection
on the oscilloscope button/menu bar, just like any other control window.
You can make an unlimited number of custom control windows, enabling
each person who uses the oscilloscope, in a shared environment, to
have their own unique control window. Since the control windows are
stored as files on the hard drive, they can easily be transferred
to other TDS5000B or TDS/CSA7000B Series oscilloscopes, or they can
even be e-mailed to a coworker around the world when the need arises.
MyScope control windows will benefit all oscilloscope users, from
eliminating the ramp-up time that many face when returning to the
lab after not using an oscilloscope for a while, to the power user
who can now operate far more efficiently. Everything you need is found
in one control window rather than having to constantly navigate through
menu after menu to repeat similar tasks.
Right Clicks
Right-mouse-click menus make simple things as they should be
– simple. Right-click menus are context sensitive, meaning the choices
presented in the menu depend on where you right clicked the mouse.
This makes right-click menus extremely intuitive. Want to change the
cursor type? Right click on a cursor or the cursor readouts. Want
to change the reference levels of an automatic measurement? Right
click on the measurement. Want to change trigger parameters? Right
click on the trigger readouts. Want to change a waveform’s color?
Right click on the waveform handle. Virtually all objects on the oscilloscope
display have right-click menus associated with them that include all
the appropriate actions or features relative to those objects. There
are also right-click menus for regions of the display in addition
to just objects. For example, right clicking in the main graticule
brings up a menu with choices such as Clear Data, Default Setup, Autoset,
Screen Captures, Save All Waveforms, and Add Screen Text, providing
single-click access to many of your most commonly performed tasks.
Pinpoint™ Triggering
Trigger on glitches down
to 100 ps wide.
Pinpoint trigger system
provides the most advanced, highest performance triggering available.
The ability to trigger an oscilloscope on events
of interest is paramount in high-speed debug and validation. Whether
you're trying to find a system error, or need to isolate a section
of a complex signal for further analysis, Tektronix’ Pinpoint triggering
provides the solution. The Pinpoint trigger system uses Silicon Germanium
(SiGe) technology to provide trigger sensitivity of up to 9 GHz (TDS6000C
models), and allows selection of all trigger types on both A and B
trigger circuits. It can capture glitches down to 100 ps wide with
1 psRMS trigger jitter typical (TDS6000C models). Other
trigger systems offer multiple trigger types only on a single event
(A Event), with delayed trigger (B Event) selection limited to edge-type
triggering and often does not provide a way to reset the trigger sequence
if the B Event doesn't occur. But Pinpoint triggering provides the
full suite of advance trigger types on both A and B triggers with
Reset triggering to begin the trigger sequence again after a specified
time, state, or transition so that even events in the most complex
signals can be captured. Other oscilloscopes typically offer less
than 20 trigger combinations; Pinpoint triggering offers over 1400
combinations, all at full performance.
Protocol Triggering
and Decoding Software (Opt. PTD)
Decode 8b/10b data streams
and set up to 4 data words (40 bits) to trigger on.
Easily decode 8b/10b and other encoded serial data streams,
and set desired encoded words for the serial pattern trigger to capture.
8b/10b decoding on data rates over 10 GB/s is possible on the TDS6000C
models. Option PTD recovers the clock signal, identifies the transitions,
and decodes the characters and other protocol data. The TDS6000C models
can trigger on up to four consecutive 10 bit words or specified error
conditions at data rates up to 3.125 Gb/s. And you can see the captured
bit sequences decoded into their words for convenient analysis.
Unparalleled Analysis
Waveform data analysis can take
many forms. Whether it’s a simple math expression, waveform mask testing,
a pass/fail compliance test, or a custom application that you develop,
the TDS6000B/C Series offer the industry's most comprehensive set
of analysis and compliance tools.
Built-in Analysis Tools
Measurement System. Enables over 50 parametric
measurements in the amplitude, time, and statistical domain.
8.5 Gb/s TDSRT-Eye™ diagram on the TDS6154C.
Waveform Math Equation Editor. Enables boundless
analysis on waveform data.
Standard tools built
into the TDS6000B/C offer a wide range of analysis capabilities including
Cursors, Measurements, Math Equation Editor, Serial Data Communications
Mask Testing (with Opt. SM), and Spectrum Analysis.
Technology-specific
Software Solutions
TDSRT-Eye™ software.
Compliance and analysis for testing high-speed serial standards. Eye
diagrams at data rates to beyond 6.25 Gb/s. Software clock recovery
to ≥10 Gb/s.
TDSJIT3 v2.0 Jitter analysis
for validation and debug of high-speed digital systems.
TDSET3 10/100/1000BaseT Compliance.
TDSDVI Digital Visual Interface Compliance.
Validation, debug, and compliance testing often require
automated test tools that quickly analyze waveform data and provide
the answers you need. The TDS6000B/C models provide complete “turn-key”
solutions for the most demanding technologies. Software options are
also available for performing validation and compliance measurements
on emerging industry standards.
OpenChoice® Analysis
Access data seamlessly for your own custom application
such as this Microsoft Excel example.
Designing
your own custom solution? The analysis and networking features of
OpenChoice software adds more flexibility to Tektronix open Windows
XP oscilloscopes:
- Fast PCI bus speeds communication between
the data acquisition processor and the Microsoft Windows desktop.
- ActiveX controls to connect the oscilloscope to popular Windows
applications - WITHOUT leaving the application.
- PnP drivers
to control the scope from LabVIEW and Lab Windows/CVI running directly
on the oscilloscope, or running on external PCs.
Support
for application development environments includes Visual BASIC, .NET,
C, C++, MATLAB, LabVIEW, and LabWindows/CVI.
Differential
Probing with Performance to >12 GHz
P7313 12.5 GHz
and P7380 8 GHz Differential Probes.
Most high-speed
signals today are differential. The P7313 12.5 GHz Differential Probing
System provides true differential connection to the device under test
for a variety of connection requirements: solder-in, hand held, and
fixtured. Typical system performance with the TDS6154C exceeds 12.5 GHz
bandwidth. The P7313 offers the lowest loading, highest signal fidelity,
and lowest cost per connection in the industry. Versatile, inexpensive
Tip-Clip™ adapters provide the optimum solution for virtually any
connection need.
The P7380SMA Differential Probing System provides
a 50 Ω per side termination network with a termination voltage. The
termination voltage can be applied externally or through the TekConnect® interface from the TDS6000B/C probe menu, or not used at
all. A gain switch provides two different sensitivity settings of
the probe and an Aux Out provides an inverted version of the signal
for driving other equipment. The P7380SMA probing system used with
the TDS6000B/C oscilloscopes provides an ideal solution for validation
and compliance testing of high-speed serial data links.
Characteristics
Vertical System
|
Characteristic
|
TDS6604B
|
TDS6804B
|
TDS6124C
|
TDS6154C
|
|
Input Channels
|
4
|
|
Bandwidth, Rise Time (DSP)
|
N/A
|
8 GHz
50 ps (10-90%) (typical)
35 ps (20-80%)
(typical)
|
12 GHz
35 ps (10-90%) (typical)
24 ps (20-80%)
(typical)
|
15 GHz
28 ps (10-90%) (typical)
19 ps (20-80%)
(typical)
|
|
True Analog Bandwidth (-3 dB), typical Rise Time
|
6 GHz
70 ps (10-90%) (typical)
53 ps (20-80%)
(typical)
|
7 GHz
62 ps (10-90%) (typical)
43 ps (20-80%)
(typical)
|
12 GHz
35 ps (10-90%) (typical)
24 ps (20-80%)
(typical)
|
12 GHz
35 ps (10-90%) (typical)
24 ps (20-80%)
(typical)
|
|
Hardware Bandwidth Limits
(Requires TCA-1MEG Adapter)
|
Full, 250 MHz, or 20 MHz
|
|
Input Coupling
|
DC, GND
|
|
Input Impedance
|
50 Ω ±2.5%
|
50 Ω ±2%
|
|
Input Sensitivity, 50 Ω
|
Full-scale 50 mV to 10 V (5 mV/div*2 to 1 V/div).
Full-scale is the peak-peak digitizer range
at a given sensitivity.
V/div = Full-scale/10.
|
|
Vertical Resolution
|
8 bit (>11 bit with averaging)
|
|
Max Input Voltage, 50 Ω
(Also determined by TekConnect® accessory)
|
<1 VRMS for <100 mV/div,
<7 VRMS for ≥100 mV/div settings.
|
<1 VRMS for <1 V full-scale,
<5.5 V for ≥1 V full-scale.
|
|
DC Gain Accuracy
|
±(2.5% + (2% × offset))
|
±2%
|
|
Position Range
|
±5 divisions
|
|
Offset Range*3
|
Full-scale settings:
100 mV to 500 mV:
±0.5 V
505 mV to 995 mV: ±0.25 V
1 V to 5 V: ±5 V
5.05 V to 10 V: ±2.5 V
|
Full-scale settings:
100 mV: ±0.45 V
200 mV: ±0.4 V
500 mV: ±0.25 V
1 V: ±4.5 V
2 V: ±4.0 V
5 V: ±2.5 V
10 V: 0
General formula
for offset range:
100 mV to 995 mV: ±|0.5 V - Full-scale/2|
1 V to 10 V: ±|5 V - Full-scale/2|
|
|
Offset Accuracy
|
±(0.7% × offset + 1.5 mV + 0.1 × V/div setting)
for ranges <100 mV/div
±(0.8% × offset + 15 mV + 0.1
× V/div setting)
for ranges ≥100 mV/div
|
±(0.35% × offset + 1.5 mV + 0.1 × V/div setting)
for ranges <100 mV/div
±(0.35% × offset + 15 mV + 0.1
× V/div setting)
for ranges ≥100 mV/div
|
|
Channel-to-Channel Isolation for Any Two Channels at Equal
Vertical Scale
|
≥80:1 at 1.5 GHz
≥15:1 at rated bandwidth
|
≥150:1 at 0 to 10 GHz
>80:1 at 10 GHz to
12 GHz
≥50:1 at 12 GHz to 15 GHz
|
*2 5 mV/div is a software zoom with 7
bits digitizer resolution at 50 mV full scale.
*3 Offset range in addition to ±5 division position range.
Vertical System (Cont.)
|
Characteristic
|
Full-scale Gain Setting
|
TDS6000B
DSP OFF
|
TDS6154C
DSP ON
|
TDS6124C
DSP ON
|
BOTH
DSP OFF
|
|
Noise, typical
|
100 mV
|
950 μV
|
690 μV
|
570 μV
|
670 μV
|
|
160 mV
|
–
|
950 μV
|
840 μV
|
950 μV
|
|
200 mV
|
1.6 mV
|
1.1 mV
|
940 μV
|
1.0 mV
|
|
300 mV
|
–
|
1.6 mV
|
1.45 mV
|
1.6 mV
|
|
400 mV
|
–
|
2.1 mV
|
1.85 mV
|
2.1 mV
|
|
500 mV
|
3.55 mV
|
2.5 mV
|
2.3 mV
|
2.4 mV
|
|
800 mV
|
–
|
4.4 mV
|
3.8 mV
|
4.1 mV
|
|
900 mV
|
–
|
4.8 mV
|
4.3 mV
|
4.6 mV
|
|
1 V
|
9.5 mV
|
6.9 mV
|
5.7 mV
|
6.8 mV
|
|
2 V
|
16 mV
|
10.5 mV
|
9.5 mV
|
10.0 mV
|
|
5 V
|
35.5 mV
|
25 mV
|
23 mV
|
24 mV
|
|
10 V
|
68 mV
|
56 mV
|
46 mV
|
50 mV
|
Timebase System
|
Characteristic
|
TDS6000B Models
|
TDS6000C Models
|
|
Timebase Range
|
25 ps to 40 s/div; Interpolation down to 500 fs/pt.
|
|
Timebase Delay Time Range
|
5 ns to 250 s
|
|
Channel-to-Channel Deskew Range
|
±75 ns
|
|
Trigger Jitter (RMS)
|
<1.5 psRMS (typical)
|
<1 psRMS (typical)
|
|
Long-term Sample Rate and Delay Time Accuracy
|
<2 ppm over any ≥100 ms interval
|
|
Clock Stability
|
<1 ppm (typical)
<2.5 ppm (guaranteed)
|
|
Jitter Noise Floor
|
|
420 fsRMS (typical) over 10 µs duration or less
|
|
Delta Time Accuracy
|
[(0.06/sample rate) + (2.5 ppm × [Reading])]RMS, typical*4
|
635 fsRMS over <100 ns duration, typical*5
|
*4 Test conditions: Sample mode, at 20 GS/s.
*5 Test conditions: Sample mode, 500 mV full-scale
(50 mV/div) setting, input signal 350 mV with rise time (10% to 90%)
<50 ps.
Acquisition System
|
Characteristic
|
TDS6000B Models
|
TDS6000C Models
|
|
Max Sample Rate
|
20 GS/s on 4 channels
|
40 GS/s on 2 channels,
20 GS/s on 4 channels
|
|
Equivalent Time Sample Rate (max)
|
2 TS/s
|
|
Maximum Record Length per Channel
|
32 M (requires Opt. 4M)
|
64 M on two channels;
32 M on all 4 channels (requires
Opt. 4M)
|
|
Standard
|
2 M on all 4 channels
|
4 M on two channels;
2 M on all 4 channels
|
|
with Memory Opt. 2M
|
8 M on all 4 channels
|
16 M on two channels;
8 M on all 4 channels
|
|
with Memory Opt. 3M
|
16 M on all 4 channels
|
32 M on two channels;
16 M on all 4 channels
|
|
with Memory Opt. 4M
|
32 M on all 4 channels
|
64 M on two channels;
32 M on all 4 channels
|
Maximum Time Duration Captured at Highest Real-time
Resolution (All Channels)
|
Characteristic
|
TDS6000B Models
(All Channels)
|
TDS6000C Models
(Two Channels)
|
|
Max Duration with Standard Memory
|
100 μs
|
100 μs
|
|
Max Duration with Opt. 2M
|
400 μs
|
400 μs
|
|
Max Duration with Opt. 3M
|
800 μs
|
800 μs
|
|
Max Duration with Opt. 4M
|
1.6 ms
|
1.6 ms
|
Acquisition Modes
|
Mode
|
All Models
|
|
Waveform Database
|
Accumulate Waveform Database providing three-dimensional array
of amplitude, time, and counts
|
|
Sample
|
Acquire sampled values
|
|
Peak Detect
|
Captures narrow glitches at all real-time sampling rates
|
|
Minimum Peak Detect Pulse Width
|
50 ps
|
|
Averaging
|
From 2 to 10,000 waveforms included in average
|
|
Envelope
|
From 2 to 2×109 waveforms included in min-max envelope
|
|
Hi-Res
|
Real-time boxcar averaging reduces random noise and increases
resolution
|
|
FastFrame™ Acquisition
|
Acquisition memory divided into segments; maximum trigger
rate >310,000 waveforms per second. Time of arrival recorded with
each event
|
|
Roll Mode
|
200 kS/s at 8 M record length and 500 kS/s at 4 M record Length
on all channels.
|
Pinpoint™ Trigger System
|
Characteristic
|
All Models
|
|
Sensitivity
|
|
Internal DC Coupled
|
0.5 div DC to 50 MHz increasing to 2.5 div at 7 GHz (TDS6000B
models, typical).
0.4 div DC to 50 MHz , increasing to 1 div
at 3 GHz, increasing to 3 div at 9 GHz (TDS6000C models, typical).
|
|
External (auxiliary input)
|
250 mVDC to 50 MHz, increasing to 350 mV at 1 GHz, increasing
to 500 mV at 2 GHz (typical).
|
|
Trigger Characteristics
|
|
Main Trigger Modes
|
Auto, Normal, and Single.
|
|
A Event and Delayed B Event Trigger Types
|
Edge, Glitch, Runt, Width, Transition Time, Timeout, Pattern,
State, Setup/Hold, Window – all except Edge, Pattern, and State can
be Logic State qualified by up to two channels.
|
|
Trigger Sequences
|
Main, Delayed by Time, Delayed by Events, Reset by Time, Reset
by State, Reset by Transition.
All sequences can include separate
horizontal delay after the trigger event to position the acquisition
window in time.
|
|
Communications-related Triggers
(requires Opt. SM)
|
Support for AMI, HDB3, BnZS, CMI, MLT3, and NRZ encoded communications
signals up to 3.125 Gb/s. Select among isolated positive or negative
one, zero pulse form, or eye patterns as applicable to standard.
|
|
Serial Pattern Trigger
(requires Opt. ST)
|
64 bit serial word recognizer, bits specified in binary (high,
low, don't care) or hex format. Trigger on NRZ-encoded data up to
1.25 Gbaud. TDS6000C models only: trigger on 8b/10b encoded data up
to 3.125 Gbaud (40 bits).
|
|
Trigger Level Range
|
|
Internal
|
±12 divisions from center of screen.
|
|
External (auxiliary in)
|
±5 V.
|
|
Line
|
Fixed at 0 V.
|
|
Trigger Coupling
|
DC, AC (attenuate <60 Hz).
HF reject (attenuate
>30 kHz).
LF reject (attenuates <80 kHz).
Noise reject
(reduce sensitivity).
|
|
Trigger Holdoff Range
|
250 ns minimum to 12 s maximum.
|
Trigger Types
Edge - Positive and/or
negative slope on any channel or front-panel auxiliary input. Coupling
includes DC, AC, noise reject, HF reject, and LF reject.
Glitch - Trigger on or reject glitches of positive, negative,
or either polarity. Minimum glitch width is down to 100 ps with re-arm
time of 250 ps.
Width - Trigger on width of positive
or negative pulse (down to 100 ps) either within or out of selectable
time limits.
Runt - Trigger on a pulse that crosses
one threshold but fails to cross a second threshold before crossing
the first again. Optional time qualification.
Timeout - Trigger on an event which remains high, low, or either, for a specified
time period, selectable from 360 ps to 1 s.
Transition
- Trigger on pulse edge rates that are faster or slower than specified.
Slope may be positive, negative, or either.
Setup/Hold
- Trigger on violations of both setup time and hold time between
clock and data present on any two input channels.
Pattern
- Trigger when pattern goes false or stays true for specified
period of time. Pattern (AND, OR, NAND, NOR) specified for four input
channels defined as HIGH, LOW, or Don't Care.
State - Any logical pattern of channels (1, 2, 3) clocked by edge on channel
4. Trigger on rising or falling clock edge.
Window - Trigger on an event that enters or exits a window defined by two
user-adjustable thresholds. Event can be time or logic qualified.
Trigger Delay by Time - 5 ns to 250 seconds.
Trigger Delay by Events - 1 to 10,000,000 events.
Waveform
Measurements
Amplitude - Amplitude, High, Low, Maximum,
Minimum, Peak-to-Peak, Mean, Cycle Mean, RMS, Cycle RMS, Positive
Overshoot, Negative Overshoot.
Time - Rise time, Fall
time, Positive Width, Negative Width, Positive Duty Cycle, Negative
Duty Cycle, Period, Frequency, Delay.
Combination - Area, Cycle Area, Phase, Burst Width.
Histogram-related
- Waveform count, Hits in box, Peak hits, Median, Maximum, Minimum,
Peak-to-Peak, Mean (μ), Standard Deviation (σ), μ + 1σ, μ + 2σ, μ
+ 3σ.
Eye-pattern-related - Extinction Ratio (absolute,
%, and dB), Eye Height, Eye Top, Eye Base, Eye Width, Crossing %,
Jitter (peak-peak, RMS, and 6sigma), Noise (peak-peak and RMS), S/N
ratio, Cycle Distortion, Q-factor.
Waveform Processing/Math
Algebraic Expressions - Define extensive algebraic expressions
including waveforms, scalars, and results of parametric measurements
e.g. (Integral (CH.1-Mean(CH.1))×1.414).
Arithmetic - Add, subtract, multiply, divide waveforms and scalars.
Relational - Boolean result of comparison >, <, ≥, ≤, =, ≠.
Calculus - Integrate, differentiate.
Frequency
Domain Functions - Spectral magnitude and phase, real and imaginary
spectra.
Vertical Units -
Magnitude: Linear, dB,
dBm.
Phase: Degrees, Radians.
Window Functions
- Rectangular, Hamming, Hanning, Kaiser-Bessel, Blackman-Harris,
Gaussian, Flattop2, Tek Exponential.
Waveform Definition
- As arbitrary math expressions.
Display Characteristics
Display Type - Liquid-crystal active-matrix color display.
Display Size - Diagonal: 264 mm (10.4 in.).
Display Resolution - 1024 horizontal × 768 vertical pixels.
Waveform Styles - Vectors, Dots, Variable Persistence,
Infinite Persistence.
Computer System and Peripherals
CPU - Intel Pentium 4 processor, 2.8 GHz.
PC System RAM - 1 GB (266 MHz DDR).
Hard Disk Drive
- 40 GB removable hard disk drive: rear-panel standard or front-panel
(Opt. FHD).
USB Ports - One on front panel, four on
rear panel, USB 2.0 compliant.
CD-R/W Drive - Front-panel
CD-R/W standard or rear-panel (Opt. FHD).
Mouse - Optical
scroll-wheel model included, USB interface.
Keyboard -
Small keyboard included (fits in pouch).
Order 119-6297-xx
for full-size keyboard; USB interface and hub.
Input/Output
Ports
Front Panel
Fast Edge Output - Front-panel SMA connector provides fast edge signal.
TDS6000B models:
Amplitude 350 mV ±20% into a ≥50 Ω load, frequency 1 kHz ±5%, 200 ps
typical rise time.
TDS6000C models: Amplitude 440 mVpk-pk, ±20% in a 50 Ω load; VOH +0.3 V, VOL - 0.14 V; frequency 1 kHz
±5%, 200 ps typical rise time.
Recovered Clock - SMA connector, ≤1.25 Gb/s, Output swing ≥130 mVpk-pk into
50 Ω. Requires Opt. SM or Opt. ST to enable.
Recovered
Data - SMA connector, ≤1.25 Gb/s, Output swing of 1010 repeating
pattern 200 mV into 50 Ω. Requires Opt. SM or Opt. ST to enable.
DC Probe Calibration Output - BNC connector, ±10 VDC
for DC probe calibration. (Signal available only during probe calibration.)
AUX Trigger Output - BNC connector, provides a TTL-compatible,
polarity switchable pulse when the oscilloscope triggers.
USB 2.0 Port - One in front. Allows connection or disconnection
of USB keyboard, mouse, or storage device while oscilloscope is on.
Rear Panel
External Timebase Reference In
- BNC connector; allows timebase system to phase-lock to external
10 MHz reference.
Timebase Reference Out - BNC connector;
provides TTL-compatible output of internal 10 MHz reference oscillator.
Aux Trigger Input - BNC, see Ext Trigger specification.
Parallel Port - IEEE 1284, DB-25 connector.
Audio Ports - Miniature phone jacks for stereo microphone input
and stereo line output.
USB 2.0 Ports - Four in back.
Allows connection or disconnection of USB keyboard, mouse, or storage
device while oscilloscope power is on.
Keyboard Port - PS-2 compatible.
Mouse Port - PS-2 compatible.
LAN Port - RJ-45 connector, supports 10BaseT, 100BaseT,
and 1000BaseT.
Serial Port - DB-9 COM1 port.
Windows Video Port - 15-pin D-Sub connector on the rear panel;
connect a second monitor to use dual-monitor display mode allowing
analysis results and plots to be viewed along with the oscilloscope
display. Video is DDC2B compliant.
GPIB Port - IEEE
488.2 standard.
Scope XGA Video Port - 15-pin D-Sub
connector on the rear panel, video is IBM XGA compatible. Connect
to show the oscilloscope display, including live waveforms on an external
monitor or projector. The primary Windows desktop can also be displayed
on an external monitor using this port.
Power Source
Power - 100 to 240 VRMS, ±10%, 50/60 Hz; 115 VRMS ±10%, 400 Hz; CAT II, <500 W typical (650 VA).
Option SM
156 Standards Masks Supported -
ITU-T (1.544 Mb/s - 155 Mb/s)
ANSI T1.102 (1.544 Mb/s
- 155 Mb/s)
Ethernet IEEE Std 802.3, ANSI X3.263 (1.544 Mb/s
- 3.125 Gb/s XAUI)
Sonet/SDH (51.84 Mb/s - 2.4883 Gb/s)
Fibre Channel (133 Mb/s - 4.25 Gb/s*6)
InfiniBand
(2.5 Gb/s)
USB (12 Mb/s - 480 Mb/s)
Serial ATA (1.5 Gb/s,
3.0 Gb/s)
Serial Attached SCSI (1.5 Gb/s, 3.0 Gb/s)
IEEE 1394b (491.5 Mb/s - 1.966 Gb/s)
Rapid I/O (1.25 Gb/s
- 3.125 Gb/s)
OIF Standards (2.488 Gb/s - 3.11 Gb/s)
PCI Express (2.5 Gb/s)
*6 4.25 Gb/s mask supported
using Glitch Trigger.
Physical Characteristics
|
Benchtop Configuration
|
|
Dimensions
|
mm
|
in.
|
|
Height
|
282
|
11.1
|
|
Width
|
457
|
18.0
|
|
Depth
|
540
|
23.25
|
|
Weight
|
kg
|
lb.
|
|
Net
|
21
|
47
|
|
Shipping
|
32.3
|
71.2
|
|
Rackmount Configuration
|
|
Dimensions
|
mm
|
in.
|
|
Height
|
267
|
10.5
|
|
Width
|
491
|
19.3
|
|
Depth
|
638
|
25.1
|
|
Weight
|
kg
|
lb.
|
|
Net
|
22
|
49
|
|
Kit
|
5.6
|
12.25
|
Mechanical
|
Cooling -
Required Clearance
|
mm
|
in.
|
|
Top
|
0 or >76
|
0 or >3
|
|
Bottom
|
25
|
1
|
|
Left side
|
76
|
3
|
|
Right side
|
76
|
3
|
|
Front
|
0
|
0
|
|
Rear
|
0
|
0
|
Environmental
Temperature
Operating -
Excluding CD-R/W drive:
TDS6000B models:
+5 °C to +40 °C.
TDS6000C models: +10 ºC to +45 ºC.
Including CD-R/W drive: +10 °C to +40 °C.
Nonoperating
- -22 °C to +60 °C.
Humidity
Operating
- 20% to 80% relative humidity with a maximum wet-bulb temperature
of +29 °C at or below +50 °C, noncondensing. Upper limit derated to
25% relative humidity at +50 °C.
Nonoperating - 5%
to 90% relative humidity with a maximum wet-bulb temperature of +29 °C
at or below +60 °C, noncondensing. Upper limit derated to 20% relative
humidity at +60 °C.
Altitude
Operating
- 10,000 ft. (3,048 m).
Nonoperating - 40,000 ft.
(12,190 m).
Random Vibration
Operating
- 0.000125 g2/Hz from 5 to 350 Hz, -3 dB/octave from
350 to 500 Hz, 0.0000876 g2/Hz at 500 Hz. Overall level
of 0.24 gRMS.
Nonoperating - 0.0175 g2/Hz from 5 to 100 Hz, -3 dB/octave from 100 to 200 Hz, 0.00875
g2/Hz from 200 to 350 Hz, -3 dB/octave from 350 to 500 Hz,
0.006132 g2/Hz at 500 Hz. Overall level of 2.28 gRMS.
Certifications
Electromagnetic Compatibility
- 89/336/EEC.
Safety - UL 3111-1, CSA1010.1, EN61010-1,
IEC 61010-1.
Ordering Information
TDS6154C
15 GHz
Digital Storage Oscilloscope.
TDS6124C
12 GHz Digital
Storage Oscilloscope.
TDS6804B
8 GHz Digital Storage
Oscilloscope.
TDS6604B
6 GHz Digital Storage Oscilloscope.
TDS6000B and TDS6000C Models Include: Accessory pouch,
front cover, USB mouse, USB mini-keyboard, user manual (071-1658-xx),
Documentation CD-ROM (063-3760-xx), product software CD-ROM (020-2572-xx
for TDS6000B models; 020-2647-xx for TDS6000C models), operating system
restoration CD-ROM (020-2569-xx), optional applications software CD-ROM
(020-2450-xx), performance verification procedure PDF file, GPIB programmer's
reference (on product software CD-ROM), calibration certificate documenting
NIST traceability, 2 540-1 compliance and ISO9000, power cord, one-year
warranty. (4) TekConnect® to SMA adapters (TCA-SMA) , Deskew
Fixture (067-0484-xx).
Please specify power plug when ordering.
Instrument Options
Power Plug Options
Opt. A0 - North America power.
Opt. A1 - Universal EURO power.
Opt. A2 - United Kingdom power.
Opt. A3 - Australia power.
Opt. A5 - Switzerland
power.
Opt. A6 - Japan power.
Opt. A10 - China power.
Opt. A99 - No power cord or AC adapter.
Manual Language Options
Opt. L5 - Japanese
manual
Opt. L7 - Simplified Chinese manual
Opt. L10 - Russian manual
Disk Drive Options
Opt. FHD - Front-panel 40 GB removable hard disk drive,
replaces front-panel CD-R/W which is moved to the back.
Cables
GPIB Cable (1 m) - Order 012-0991-01.
GPIB Cable (2 m) - Order 012-0991-xx.
RS-232
Cable - Order 012-1298-xx.
Centronics Cable - Order
012-1214-xx.
Mounting Options
Opt. 1K - K4000 Scope Cart.
Opt. 1R - Rackmount Kit.
Service Options
Opt. C3 - Calibration Service
3 Years.
Opt. C5 - Calibration Service 5 Years.
Opt. D1 - Calibration Data Report.
Opt. D3
- Calibration Data Report 3 Years (requires Opt. C3).
Opt. D5 - Calibration Data Report 5 Years (with Opt. C5).
Opt. R3 - Repair Service 3 Years.
Opt. R5
- Repair Service 5 Years.
Acquisition Memory Options
TDS6000B Models:
Opt. 2M - 8 MSamples
on all channels.
Opt. 3M - 16 MSamples on all channels.
Opt. 4M - 32 MSamples on all channels.
TDS6000C
Models:
Opt. 2M - 16 MSamples on two channels;
8 MSamples on all channels.
Opt. 3M - 32 MSamples on
two channels; 16 MSamples on all channels.
Opt. 4M - 64 MSamples on two channels; 32 MSamples on all channels.
Software Options
Opt. CP2*7 - TDSCPM2 ANSI/ITU Telecom Pulse Compliance Testing Software (requires
Opt. SM).
Opt. DVI - TDSDVI DVI Compliance Test Solution.
Opt. ET3 - TDSET3 Ethernet Compliance Test Software.
Opt. HT3 - HDMI Compliance Test Software.
Opt. JA3 - TDSJIT3 v2.0 Advanced Jitter Analysis Software.
Opt. JE3 - TDSJIT3 v2.0 Essentials Jitter Analysis Software.
Opt. J2 - TDSDDM2 Disk Drive Analysis Software.
Opt. PW3*8 - TDSPWR3 Power Measurement and Analysis
Software.
Opt. RTE - TDSRT-Eye™ Serial data compliance
and analysis software.
Opt. RJA - Software Bundle of
TDSJIT3 v2.0 (Opt. J3A), TDSRT-Eye (Opt. RTE), and Protocol Trigger
Decode (Opt. PTD).
Opt. RSA - Software Bundle of TDSJIT3
v2.0 (Opt. J3A), RT-Eye (Opt. RTE), Serial Trigger (Opt. ST), Protocol
Trigger and Decode (Opt. PTD), and Serial Mask testing (Opt. SM).
Opt. PCE*9 - PCI Express™ Compliance Module
for Opt. RTE.
Opt. PTD - Protocol Trigger and Decode
(provides protocol decode on all models, requires Opt. ST for protocol
triggering on the TDS6000C Series only).
Opt. IBA*9 - InfiniBand® Compliance Module for Opt. RTE.
Opt. SM - Serial Communications Mask Testing (enables
hardware clock recovery).
Opt. SST - SATA and SAS Analysis
Software Module for Opt. RTE.
Opt. ST - Serial Pattern
Trigger (enables hardware clock recovery)
Opt. USB*10 - TDSUSBS USB 2.0 Compliance Test SW Only.
*7 Requires Opt. SM.
*8 Opt. 3M and a TCA-1MEG
TekConnect 1 MΩ buffer amplifier are recommended for use with this
software.
*9 Requires Opt. RTE (Serial Data Compliance
and Analysis).
*10 Requires TDSUSBF (USB Test Fixture).
Recommended Accessories
P7313 - 12.5 GHz Differential
Probe.
P7380 - 8 GHz Differential Probe.
P7380SMA - 8 GHz Differential Probing System.
AFTDS
- Telecom Differential Electrical Interface Adapter (for line
rates < 8 MB/sec; requires TCA-BNC adapter).
Keyboard
(USB interface) Full-size Keyboard with 4-port USB hub - Order
119-6297-xx.
Transit Case - Order 016-1942-xx.
TekConnect® Adapters -
TCA-1MEG: TekConnect® High-impedance Buffer Amplifier. Includes P6139A Passive
Probe.
TCA-SMA: TekConnect-to-SMA Adapter.
TCA-N: TekConnect-to-N
Adapter.
TCA-BNC: TekConnect-to-BNC Adapter.
TCA75:
4 GHz precision TekConnect® 75 Ω to 50 Ω Adapter with 75 Ω
BNC Input Connector.
Oscilloscope Cart - Order
K4000.
Software -
WSTRO: WaveStar™ Waveform Capture
and Documentation Software.
Test Fixtures -
TDSUSBF: TDSUSB Test Fixture for use with Opt. USB.
After Purchase Upgrades of TDS6000B and TDS6000C Series Oscilloscopes
|
Acquisition Record Length
|
|
Current Record Length
|
Desired
|
Order
|
|
To upgrade your oscilloscope, order option as
noted.
|
|
Standard
|
8 MSamples (16 MSamples TDS6000C)
|
TDS6BUP Opt. M02
|
|
16 MSamples (32 MSamples TDS6000C)
|
TDS6BUP Opt. M03
|
|
32 MSamples (64 MSamples TDS6000C)
|
TDS6BUP Opt. M04
|
|
8 MSamples
|
16 MSamples (32 MSamples TDS6000C)
|
TDS6BUP Opt. M23
|
|
32 MSamples (64 MSamples TDS6000C)
|
TDS6BUP Opt. M24
|
|
16 MSamples
|
32 MSamples (64 MSamples TDS6000C)
|
TDS6BUP Opt. M34
|
|
Order
|
Description
|
|
TDS6BUP Opt. SM
|
Serial Comm Mask Testing
|
|
TDS6BUP Opt. ST
|
Serial Pattern Trigger
|
|
Included in Opt. SM and ST
|
Hardware Clock Recovery for Serial Data Stream
|
|
Analysis and Compliance Software
|
|
TDS6BUP Opt. CP2 (Requires Opt. SM)
|
ANSI/ITU Telecom Pulse Compliance
|
|
TDS6BUP Opt. DVI
|
DVI Compliance Test
|
|
TDS6BUP Opt. ET3
|
Ethernet Compliance
|
|
TDS6BUP Opt. JA3
|
JIT3 V2.0 Advanced Jitter Analysis Software
|
|
TDS6BUP Opt JE3
|
JIT3 V2.0 Essentials Jitter Analysis Software
|
|
TDS6BUP Opt. J2
|
DDM2 Disk Drive Analysis Software
|
|
TDS6BUP Opt. PW3
|
Power Measurement/Analysis
|
|
TDS6BUP Opt PTD
|
Protocol Trigger and Decode (trigger capability for TDS6000C
only, requires Opt. ST)
|
|
TDS6BUP Opt RTE
|
RT-Eye Serial Data Compliance and Analysis Software
|
|
TDS6BUP Opt PCE (Requires Opt. RTE or RJA or RSA)
|
PCI Express™ Compliance Module for Opt. RTE
|
|
TDS6BUP Opt IBA (Requires Opt. RTE or RJA or RSA)
|
InfiniBand® Compliance Module for Opt. RTE
|
|
TDS6BUP Opt. USB (Requires TDSUSBF test fixture)
|
USB 2.0 Compliance, Software Only
|
|
TDS6BUP Opt. RJA
|
RT-Eye Serial Data Analysis Software, TDSJIT3 v2.0 Advanced
Jitter Analysis, Opt. PTD Protocol Trigger and Decode Software Bundle
|
|
TDS6BUP Opt. RSA
|
RT-Eye Serial Data Analysis, TDSJIT3 v2.0 Advanced Jitter
Analysis, Opt. SM Serial Mask, Opt. ST Serial Trigger, Opt. PTD Protocol
Trigger and Decode Software Bundle
|