Features
& Benefits
- High-resolution DC Parametric Measurements With
the 370B (Resolution Down to 1 pA or 2 µV)
- High Voltage and Current
Sourcing With the 371B (Up to 3,000 V or 400 A)
- Built-in Cursor Measurements
- Dot, Window and Function Line
- Kelvin Sense Measurements
- Sweep
Measurement Mode
- Waveform Comparison and Averaging
- Fully
Programmable
- 1.44 MB Floppy Drive Stores Setup and Bitmap of Curves
- Direct
Hardcopy with Third-party Printer
Applications
- Parametric
Characterization of Semiconductors
- Failure Analysis
- Data
Sheet Generation
- Manufacturing Test
- Process Monitoring and
Quality Control
- Incoming Inspection
- Component Matching
370B and
371B Programmable Curve Tracers
Tektronix curve tracers offer power
and versatility to test the DC characteristics of semiconductor devices. The
370B and 371B curve tracers combine a simple-to-use front panel, digital acquisition
and display, and programmability to serve a variety of application needs.
In
the R&D lab, Tektronix curve tracers are used for characterization of
new designs, extraction of SPICE parameters, failure analysis and data sheet
generation. In manufacturing test, Tektronix curve tracers are used to verify
device quality and process monitoring. They are also used for incoming inspection
to verify device performance, perform failure analysis and match components.
The
World Standard for High-resolution Curve Tracers
370B.
The
370B, the world standard for high-resolution curve tracers, provides up to
20 A/2,000 V sourcing capability combined with 1 pA and 50 µV measurement
resolution. The 370B performs DC parametric characterization of integrated
circuits, transistors, thyristors, diodes, SCRs, MOSFETs, electro-optic components,
solar cells, solid-state relays and other semiconductor devices. It has pushbutton
source and measurement configuration so it's easy to change from one test
to the next.
The Industry Leader in High-power Curve Tracers
371B.
The
371B, the industry leader in high-power curve tracers, performs DC parametric
characterization on a wide variety of power semiconductors including thyristors,
SCRs, IGBTs and power MOSFETs. The high-voltage collector mode permits testing
the Off-Characteristics of a device up to 3,000 volts. The pulsed high-current
collector mode provides output current pulses greater than 400 amps peak to
test On-Characteristics and permits high-power testing up to 3,000 watts.
In the sweep measurement mode, the 371B automatically constructs a family
of curves while stimulating the device with low-duty-cycle pulses. With this
capability, power curves can be displayed without excessive heating of the
device.
Interactive Programmable Control
Interactive control
of all 370B and 371B measurements is accomplished from the full-featured front
panel or over the GPIB. Every operating parameter can be controlled using
the GPIB controller. 370B and 371B Series LabView drivers are available at
the Tektronix.com web site. These
drivers have many of the building blocks to create a custom measurement solution.
Store
and Recall Setups and Digitized Curves
Up to 80 digitized characteristic
curves can be stored on a diskette or in internal non-volatile memory and
recalled at the touch of a button. A live curve can then be compared with
a previously stored curve to assess temperature drift or other changes in
operating parameters. To help identify the data, up to 24 characters of text
may be used to label or annotate the curve data.
Operating parameters
can be adjusted, stored and recalled using several storage methods including
the 370B and 371B system memory, the built-in 1.44 MB floppy drive, or sent
externally via the GPIB controller.
Automated Cursor
The 370B
and 371B provide three cursor measurement modes. The dot cursor provides direct
screen readout of voltage, current, gm or DC beta at any point. The window
cursor can be positioned between two curves to measure small signal beta or
gm and can also be used for visual go/no-go tests. The function line cursor
provides screen readout of a slope or intercept value.
Sweep Measurement
Mode
In the sweep measurement mode, the curve tracer automatically
constructs a family of curves. The 371B achieves this with low duty-cycle
pulses. With this capability, power curves can be displayed without excessive
heating of the device.
Direct Hardcopy Printouts
Curve output
data can be sent directly from the 370B and 371B to a Citizen iDP-3240thermal
printer. Printing can continue while the 370B and 371B perform the next tasks.
Characteristics
370B
Curve Tracer Characteristics
Collector Supply
|
Peak Voltage (Range)
|
16 V
|
80 V
|
400 V
|
2,000 V
|
|
Polarity Modes
|
AC, ± DC, ± Leakage, ± Rectified Sine
|
|
Available Current
|
|
Peak Current
|
10 A
|
2 A
|
0.4 A
|
0.05 A
|
|
Peak Current (PULSE mode)
|
20 A
|
4 A
|
0.8 A
|
0.1 A
|
|
Available Resistance
|
|
Minimum Series Resistance
|
0.26 Ω
|
6.4 Ω
|
160 Ω
|
20 kΩ
|
|
Maximum Series Resistance
|
800 Ω
|
20 kΩ
|
500 kΩ
|
12.5 MΩ
|
Available Peak Power
16, 80 and 400 V
Range - 0.08, 0.4, 2, 10, 50, 220 W.
2,000 V Range - 0.08,
0.4, 2, 10, 50 W.
Variable Collector Supply Settings As a % of
Maximum Peak Voltage - 0 to 100% in 0.1% increments.
Looping
Compensation (typical) Range - at least 100 pF.
Step Generator
Accuracy
Incremental
- 1.5%.
Absolute - Less than 1.5% of total output + 3%
x STEP AMPLITUDE setting + 1 mV or 1 nA.
Offset Control Range - Variable
from -10 to +10 times STEP AMPLITUDE.
Resolution - STEP AMPLITUDE
setting x 1%.
Current Mode
Amplitude Range - 50 nA
to 200 mA in a 1-2-5 sequence of 21 steps.
Maximum Current - 20x
STEP AMPLITUDE setting, 10x STEP AMPLITUDE setting when control is set to
200 mA.
Maximum Voltage - At least 10 V.
Maximum
Opposing Offset Current - 10x STEP AMPLITUDE.
Maximum Opposing
Volts - Less than 15 V.
Ripple Plus Noise - Less than 0.5%
x STEP AMPLITUDE + 10 nA.
Voltage Mode
Amplitude
Switch Range - 50 mV to 2 V, in a 1-2-5 sequence of 6 steps.
Maximum
Voltage - 20x STEP AMPLITUDE.
Maximum Current - At least
500 mA at 10 V or less, at least 200 mA at 15 V, at least 10 mA at 40 V.
Short
Circuit Current Limiting - 500 mA, +50%, -20%.
Maximum Opposing
Offset Current - 10x STEP AMPLITUDE.
Maximum Opposing Current
- Less than 20 mA.
Ripple Plus Noise - Less than 0.5% x
STEP AMPLITUDE + 10 mV.
Step Rates - 2x line frequency (1 x
line frequency in AC collector supply mode). Steps occur at 0 collector voltage.
Pulsed
Steps - 80 µs or 300 µs wide, ±10%.
Steps and Offset Polarity
- Corresponds to Collector Supply Polarity when STEP GENERATOR POLARITY
INVERT is disabled. Opposite to Collector Supply Polarity when STEP GENERATOR
POLARITY INVERT is selected or CONFIGURATION switch is set to BASE COMMON.
BASE COMMON configuration disables STEP GENERATOR POLARITY INVERT.
Number
of Steps - Ranges from 0 to 10.
AUX Supply
Range
- From -40 to +40 volts, with 20 mV step resolution.
Accuracy
(typical) - Within 50 mV + 1.5% of total output.
Output Current
- At least 100 nA at ±20 V.
At least 10 mA at ±40 V.
Ripple
Plus Noise - Less than 50 mVp-p.
Non-store Vertical
Deflection System
Collector/Base Current
Range
- 1 µA/div to 2 A/div in a 1-2-5 sequence.
X10 MAG extends maximum sensitivity
to 100 nA/div (1 nA resolution).
Accuracy (typical) - Within
2% of WINDOW cursor readout + 0.1 x VERT/DIV setting.
Emitter Current
Range
- 1 nA/div to 2 mA/div in a 1-2-5 sequence X10 MAG extends maximum sensitivity
to 100 pA/div.
Accuracy (typical) - Within 2% of WINDOW cursor
readout + 0.1 x VERT/DIV settings + 1 nA.
Maximum Displayed Noise
or Ripple - Depending on setting of MAX PEAK VOLTS.
|
Max. Volts Range
|
16 V
|
80 V
|
400 V
|
2000 V
|
|
Noise or Ripple
|
1 nAp-p
|
1 nAp-p
|
2 nAp-p
|
5 nAp-p
|
Step Generator Display
Range - 1 step/division.
1 step/10 divisions
with X10 MAG.
10 steps/division with STEP MULTI .1x.
Accuracy
- Within 0.3 division.
Display Offset (typical)
Range
- ±10 divisions with 0.1 div resolution.
Accuracy - Within
1.5% of offset + 0.1 x VERT/DIV setting.
Display X10 MAG (typical)
Accuracy
- Within 1.5% of window cursor readout + 0.3 x VERT/DIV setting.
Display
Invert (typical)
Accuracy - Within 0.1 x VERT/DIV setting.
Digital
Storage Vertical Acquisition
A/D Converter
Resolution
- 10-Bits for 10.24 divisions, 100 counts per division.
Max.
Data Points - 1024.
Max. Sampling Rate - Line frequency
x 1024.
Min. Sampling Rate - Line frequency x 2.
Collector/Base
Current
Range - 1 µA/div to 2 A/div in a 1-2-5 sequence.
X10 MAG
extends maximum sensitivity to 100 nA/div (1 nA resolution).
Accuracy
- Within 1.5% of DOT cursor readout +0.05 x VERT/DIV setting.
Emitter
Current
Range - 1 nA/div to 2 mA/div in a 1-2-5 sequence.
X10 MAG
extends max sensitivity to 100 pA/div (1 pA resolution).
Accuracy
- Within 1.5% of DOT cursor readout +0.05 x VERT/DIV setting + 1 nA.
Step
Generator Display
Range - 1 step/division, 1 step/10 divisions
with X10 MAG, 10 steps/division with STEP MULTI .1x.
Accuracy - Within
0.3 division.
Display Offset (typical)
Range - ±10 divisions
with 0.1 div resolution.
Accuracy - Within 1.5% of offset +
0.06 x VERT/DIV setting.
Display X10 MAG (typical)
Accuracy
- Within 1.5% of window cursor readout + 0.3 x VERT/DIV setting.
Display
Invert (typical)
Accuracy - Within 0.04 x VERT/DIV setting.
Non-store
Horizontal Deflection System
Collector Volts
Range
- 50 nV/div to 500 V/div in a 1-2-5 sequence.
X10 MAG extends maximum
sensitivity to 5 mV/div (50 µV resolution).
Accuracy (typical)
- Within 2% of WINDOW cursor readout + 0.1 x HORIZ/DIV setting.
Base/Emitter
Volts
Range - 50 mV to 5 V/div in a 1-2-5 sequence.
X10 MAG
extends maximum sensitivity to 5 mV/div (50 µV resolution).
Accuracy
(typical) - Within 2% of WINDOW cursor readout + 0.1 x HORIZ/DIV setting.
Step
Generator Display
Range - 1 step/division, 1 step/division
with X10 MAG, 10 steps/division with STEP MULTI .1x.
Accuracy - Within
0.3 division.
Display Offset
Range - ±10 divisions
in 0.1 division steps.
Accuracy - Within 1.5% of offset +0.1
x HORIZ/DIV setting.
Display X10 MAG
Accuracy - Within
1.5% of window cursor readout + 0.3 x HORIZ/DIV setting.
Display
Invert
Accuracy - Within 0.1 x HORIZ/DIV setting.
Digital
Storage Horizontal Acquisition
A/D Converter
Resolution
- 10 bits for 10.24 divisions, 100 counts per division.
Max.
Data Points - 1024.
Max. Sampling Rate - Line frequency
x 1024.
Min. Sampling Rate - Line frequency x 2.
Collector
Volts
Range - 50 mV/div to 500 V/div in a 1-2-5 sequence.
X10 MAG
extends maximum sensitivity to 5 mV/div (50 µV resolution).
Accuracy
- Within 1.5% of DOT cursor readout + 0.05 x HORZ/DIV setting.
Acquisition
Modes
Normal, Envelope, and Average. Envelope is Vertical or Horizontal.
Average is moving average with weight of 1/16.
Safety
Standards
Met - UL3111-1, CAN/CSA C22.2 NO. 1010.1, EN 61010-1 with second amendment.
371B
Curve Tracer Characteristics
Collector Supply
Modes
(positive and negative polarities for both modes) - High current: 250 µs
pulses with maximum peak of 30 V.
High voltage: Full rectified sine wave
with maximum peak of 3,000 V.
Polarity Modes - NPN +.
PNP -.
|
|
High Voltage Mode
|
High Current (pulsed) Mode
|
High Current (pulsed) Mode
|
|
Peak Power Watts (range)
|
30 mW
|
300 mW
|
3 W
|
30 W
|
300 W
|
3 kW
|
|
Collector Peak Current
|
0.4 mA ±20%
|
4 mA ±20%
|
4 mA ±20%
|
40 mA ±20%
|
40 A
|
400 A
|
|
Maximum Peak Voltage
|
300 V + 15% - 0%
|
300 V + 15% - 0%
|
3 kV + 10% - 0%
|
3 kV + 10% - 0%
|
30 V + 10% - 5%
|
30 V + 10% - 5%
|
Variable Collector Supply Range - Continuously variable
from 0% to 100% in 0.1% increments.
Looping Compensation (high
voltage mode) - Stray capacitance between collector terminal and ground
is compensated for up to a maximum of 100 pF.
Pulsed Collector
Supply Repetition Rate -
3 kW: 1/4 (0.25 x) line frequency.
300 W:
1/2 (0.5 x) line frequency.
Pulse Width (half amplitude) - 250
µs ±10% (more than 30% of collector supply).
Circuit Breakers - Both
the high-voltage and the high-current supplies have breakers which operate
independently. The high-current output breaker opens automatically in an over-current
condition. Both breakers can be operated manually.
Step Generator
Offset
Control Range - Variable 0 to 10 times STEP/OFFSET AMPLITUDE setting.
However, 0 to 5 times when STEP/OFFSET AMPLITUDE is 5 V or 2 A.
Number
of Steps - 0 to 10. However, 0 to 5 when STEP/OFFSET.
Step
Polarity - Positive, Negative.
Step Rate -
Pulse:
3 kW:
1/4 (0.25x) the line frequency, 300 W: 1/2 (.5X) the line frequency.
Stairstep:
300 W,
3 W, 300 mW, 30 mW: 2x the line frequency.
Current Mode
Amplitude
Range
Normal Step - 1 µA to 2 mA in 1-2-5 sequence. COLLECTOR
SUPPLY HIGH VOLTAGE MODE (30 W/3 W and 300 mW/30 mW).
Pulsed Step
- 1 mA to 2 A in 1-2-5 sequence. COLLECTOR SUPPLY HIGH CURRENT MODE (3 kW/300 W).
Maximum
Current - 20 x step amplitude, except 10 times STEP/OFFSET AMPLITUDE when
control is set to 2 A per step.
Maximum Voltage - 12 V ±30%.
Accuracy:
At least 5 V.
Ripple Plus Noise - Within 1% of STEP AMPLITUDE
+ 10 nA.
Voltage Mode
Step Amplitude Range - 200 mV
to 5 V in 1-2-5 sequence.
Short Circuit Current Limiting - 100 mA
+50%, -20%.
Maximum Voltage - 10 x STEP AMPLITUDE, with 5 V
setting.
20 x STEP AMPLITUDE, except 5 V setting.
Ripple
Plus Noise - Within 1% of STEP AMPLITUDE + 10 mV.
Vertical
Deflection System
Collector Current (lc) Range
3 kW
Peak Power - 1 A/DIV to 50 A/DIV.
300 W Peak Power - 500 mA/DIV
to 5 A/DIV.
30 W Peak Power - 100 µA/DIV to 5 mA/DIV.
3 W
Peak Power - 10 µA/DIV to 500 µA/DIV.
0.3 W Peak Power - 10 µA/DIV
to 500 µA/DIV.
0.03 W Peak Power - 1 µA/DIV to 50 µA/DIV.
Collector
Current (lc) Accuracy - Within 0.1 DIV to the Vertical graticule lines.
Collector
Current (lc) Cursor Accuracy - Within 1.5% of readout +0.1 DIV of the
CURRENT/DIV setting.
Horizontal Deflection System
Collector
Supply Volts (Vce) Range
3 kW, 300 W Peak Power - 100 mV/DIV
to 5 V/DIV in a 1-2-5 sequence.
30 W, 3 W Peak Power - 50 V/DIV
to 500 V/DIV in a 1-2-5 sequence.
300 mW, 30 mW Peak Power - 5 V/DIV
to 50 V/DIV in a 1-2-5 sequence.
Step Generator Volts (Vbe) Range
- 100 mV/DIV to 5 V/DIV in a 1-2-5 sequence.
Step Generator
Volts (Vbe) Accuracy - Within 0.1 DIV of Horizontal graticules.
Digital
Storage Acquisition
A/D Converter
Resolution
- 10 bits for 1024 division, 100 points per one division.
Tolerance
- Within ±2 counts.
Max. Data Point - 1024.
Max.
Sample Rate - 1024 x line frequency.
Min. Sample Rate - 1/8
x line frequency.
Safety
Standards Met - UL3111-1,
CAN/CSA C22.2 NO. 1010.1, EN 61010-1 with second amendment.
Ordering
Information
370B
High-Resolution Programmable Curve Tracer
370B
Standard Accessories
A1001 - Blank adapter for mounting
custom sockets.
A1002 - In-line transistor adapter.
A1005
- Axial diode lead adapter.
User Manual - Order 070-A838-00
(English).
Protective Cover - Order 337-3344-02.
Power
Cord - Order 161-0066-00.
Fuses - Order 159-0260-00 (125 V,
4 A).
Order 159-0259-00 ( 250 V, 2 A).
Utility Software
- Order 063-3341-02.
370B Recommended Accessories
Service
Manual - Order 071-A842-00.
K4000 - Instrumentcart.
Rackmount
kit - Order 016-0930-00 (Opt. 1R).
371B
High-Power Programmable
Curve Tracer
371B Standard Accessories
A1002 - In-line
transistor adapter.
User Manual - Order 070-A840-00 (English).
Power
Cord - Order 161-0066-00.
Wire Set - Order 012-1371-00.
Fuses
- Order 159-0291-00 (250 V, 1 A).
Order 159-0238-00 (250 V, 4 A).
Utility
Software - Order 063-3341-02.
371B Recommended Accessories
A1001
- Blank adapter for mounting custom sockets.
A1005 - Axial
lead adapter.
Service Manual - Order 070-A843-00.
K4000
- Instrument cart.
Rackmount Kit - Order 016-0930-00 (Opt.
1R).
Power Plug Options
Opt. A1 - Euro Plug, 220V,
50 Hz
Opt. A2 - UK Plug, 240V, 50 Hz
Opt. A3 - Australian
Plug, 240V, 50 Hz
Opt. A5 - Swiss Plug, 220V, 50 Hz
Opt.
AC - China Plug, 50 Hz
Opt. A9 - No power cord, set to
220/240 V.
Service
Opt. C3 - Calibration Service 3
Years
Opt. C5 - Calibration Service 5 Years
Opt.
D1 - Calibration Data Report
Opt. D3 - Calibration Data
Report 3 Years (with Option C3)
Opt. D5 - Calibration Data
Report 5 Years (with Option C5)
Opt. R3 - Repair Service 3
Years
Opt. R5 - Repair Service 5 Years