Passive 1x/10x Voltage Probe
P2220 • P2221
|
Features & Benefits
- 1x/10x Attenuation
- 200 MHz Bandwidth
- 1.5 m Length
- UL3111-1, CSA1010.1,
EN61010-1, EN61010-2-031, and IEC61010-2-031
Applications
- Relative Low-frequency Measurements
- Low-frequency Computer and Telecom Measurements
- Power
Supplies
- Low-frequency Amplifiers
P2220,
P2221 Passive Voltage Probes
The P2220 and P2221, 200 MHz
passive voltage probes permit selection of 1x or 10x attenuation,
using a switch located on the probe head.
The P2220 probe is
compatible with TDS200, TDS1000, TDS1000B, TDS2000, TDS2000B, and
TPS2000 Series oscilloscopes. The P2220 probe also provides floating
measurement capability up to 30 VRMS with TPS2000 Series
oscilloscopes only. (Note: Do not float with TDS2000, TDS2000B, TDS1000,
TDS1000B, or TDS200 Series oscilloscopes.) The P2221 is compatible
with MSO2000 and DPO2000 Series oscilloscopes.
Characteristics
10x Position
Bandwidth
- DC to 200 MHz
Probe Length - 1.5 m
Attenuation Ratio - 10:1
Compensation Range -
P2220: 15 to 25 pF
P2221: 10 to 25 pF
Input Capacitance -
P2220: 17.0 pF (typ)
P2221: 17.0 pF
(typ)
Input Resistance - 10 MΩ
Maximum
Input Voltage - 300 V CAT II
Maximum Voltage Between
Reference Lead and Earth Ground - 30 VRMS*1
1x Position
Bandwidth - DC to 6 MHz
Probe Length - 1.5 m
Attenuation Ratio - 1:1
Compensation Range - All oscilloscopes with 1 MΩ
input
Input Capacitance -
P2220: 110 pF (typ)
P2221: 110 pF (typ)
Input Resistance - 1 MΩ
Maximum Input Voltage - 150 V CATII
Maximum
Voltage Between Reference Lead and Earth Ground - 30 VRMS*1
*1 The float voltage must be subtracted
from the tip to earth ground voltage. For example, if the reference
lead is floated to 30 VRMS while the probe is in the 10x
position, the tip voltage to the reference lead is limited to 270 VRMS.
Ordering Information
P2220
Passive
Probe, 1x/10x, 200 MHz Passive Voltage Probe.
Includes: Retractable hook tip (013-0107-xx), 6-inch ground lead (196-3466-xx),
adjustment tool (003-1433-xx), marker band set (016-1315-xx), Instruction
Manual (071-1464-xx).*2
P2221
Passive
Probe, 1x/10x, 200 MHz Passive Voltage Probe for MSO/DPO2000 Series
oscilloscopes.
Includes: Retractable hook tip (013-0107-xx),
6-inch ground lead (196-3466-xx), adjustment tool (003-1433-xx), marker
band set (016-1315-xx), Instruction Manual (071-2335-xx).*2
*2 The accessory quantities may differ when multiple
probes are shipped with an oscilloscope/probe system.
51W-17984-1, 18-NOV-2008
|
|
 |
(WebID: 13418)
|