TekVPI™ Deskew Pulse Generator Signal Source
TEK-DPG
|
Features & Benefits
- An Easy-to-use, Plug-on Accessory
for Oscilloscopes That Feature the TekVPI Interface
- Generates
Source Signals to Enable Precise Deskew Between Current and Voltage
Probes in Power Measurement Applications
- LED Status Indicator
for Quick Identification of Generator State
- Multiple Output
Signals Enable Use with a Variety of Oscilloscopes and Probes
- Output can be Turned Off to Reduce Power Consumption when Not
in Use
- Easy Access to Oscilloscope Menu Display to Control
Settings from the Oscilloscope Front Panel
Applications
- Power Supply Design
- Measurements Requiring Accurate Timing Measurements Between Channels
TEK-DPG
Deskew Pulse Generator
The TEK-DPG Deskew Pulse Generator
is a signal-generating accessory for Tektronix instruments that feature
the TekVPI™ interface, including DPO3000 Series and DPO/MSO4000 Series
oscilloscopes.
The TEK-DPG generator provides a source signal
to the Power Measurement Deskew Fixture (Tektronix part number 067-1686-xx),
used to deskew popular Tektronix probes. The generator and deskew
fixture allow you to precisely deskew the time differences between
instrument channels of current and voltage probes, necessary when
you are using the probes in power measurement applications.
Characteristics
|
Mode
|
Amplitude
|
Frequency
|
Rise Time
|
|
(all values are Typical)
|
|
Mode 1
|
1 Vp-p (-1 V..0 V)
|
1 kHz
|
0.8 ns
|
|
Mode 2
|
2 Vp-p (-2 V..0 V)
|
1 kHz
|
1.6 ns
|
|
Mode 3
|
2 Vp-p (-1 V..+1 V)
|
1 MHz
|
2 ns
|
|
Mode 4
|
8 Vp-p (-4 V..+4 V)
|
1 MHz
|
4 ns
|
Terminates Into - 50 Ω
Expected
Load - 50 Ω
Status LED Values
|
Status LED Color
|
Description
|
|
Amber
|
The generator is powered and the output at the BNC cable is not enabled. This is the default status at power-on
|
|
Green
|
The generator is powered and the output at the BNC cable is enabled
|
|
Red
|
A fault condition may have occurred. Disconnect and reconnect
the generator to clear the fault condition
|
Recommended Output Mode
|
Probe*1
|
Operation
|
TEK-DPG
Output Mode*2
|
067-1686-xx
Input Port
|
|
The recommended output mode is determined by the
combination of oscilloscope model, voltage probe, and current probe,
according to the following table. See the user instructions for the
Power Measurement Deskew Fixture, Tektronix part number 067-1686-xx,
for deskew fixture operation.
|
|
Passive Voltage Probes
|
|
P6139A, P5050,
and Other 10x Passive Probes
|
Compensate timing with respect to current probes
|
DPO3000: Mode 2
DPO/MSO4000: Mode 1
|
A or B, depending on current probe type
|
|
Active Voltage Probes
|
|
TAP1500, TAP2500, TAP3500
TDP0500, TDP1000, TDP1500
P6243, P6245, P6246, P6247, P6248, P6250, P6251
|
Compensate timing with respect to current probes
|
DPO3000: Mode 2
DPO/MSO4000: Mode 1
|
A or B, depending on current probe type
|
|
High-voltage Probes
|
|
P5200, P5205, P5210
P6135A
|
Compensate timing with respect to current probes
|
DPO3000: Mode 2 or 4*3
DPO/MSO4000 : Mode
1 or 4*3
|
A or B, depending
on current probe type
|
|
Small Current Probes
|
|
TCP0030
TCP202, TCP312, TCP305
A6302, A6312
|
Compensate timing with respect to voltage probes
|
DPO3000: Mode 2 or 4*3
DPO/MSO4000: Mode
1 or 4*3
|
A
|
|
Large Current Probes
|
|
TCP0150
TCP303
A6303
|
Compensate timing with respect to voltage probes
|
DPO3000: Mode 4
DPO/MSO4000: Mode 4
|
B
|
*1 Note that some probe models listed
do not utilize the TekVPI interface. Probes which use the TekProbe-BNC
interface require the TPA-BNC adapter to connect to a TekVPI oscilloscope.
*2 TEK-DPG is not supported by the DPO7000 Series.
The Probe Calibration output from the DPO7000 Series generates the
appropriate source signal for the Power Measurement Deskew Fixture.
*3 Mode 4 is a universal mode for high attenuation
probes. High-voltage probes and current probes can always use Mode
4.
Physical Characteristics
|
Dimensions
|
cm
|
in.
|
|
Height
|
3.048
|
1.20
|
|
Width
|
4.064
|
1.60
|
|
Length
|
11.73
|
4.62
|
|
Cable Length
|
41.225
|
16.2
|
|
Weight
|
kg
|
lb.
|
|
Shipping
|
0.325
|
0.7175
|
Power Requirements
- TEK-DPG
is powered directly from DPO3000, DPO/MSO4000, and DPO7000 Series
Oscilloscopes, using the TekVPI probe interface.
Environmental
|
Characteristic
|
Description
|
|
Temperature
|
|
Operating
|
0 °C to +50 °C
|
|
Nonoperating
|
-40 °C to +71 °C
|
|
Humidity
|
|
Operating
|
5% to 90% relative humidity (% RH) at up to +40 °C
5% to 60% RH above +40 °C up to +50 °C, noncondensing
|
|
Nonoperating
|
5% to 90% RH (relative humidity) at up to +40 °C
5%
to 60% RH above +40 °C up to +50 °C, noncondensing
|
|
Altitude
|
|
Operating
|
Up to 3000 meters (9,843 feet)
|
|
Nonoperating
|
Up to 12,000 meters (39,370 feet)
|
|
Regulatory
|
|
Emissions Compliance
|
EN 55011, Class A
|
|
EMC Compliance
|
EN61326, Class A
|
|
Compliance Labeling
|
WEEE (European Union)
|
Standard Warranty
One-year parts
and labor.
Recommended Oscilloscopes
- DPO3000 and DPO/MSO4000 Series Oscilloscopes with TekVPI probe interface.
- TEK-DPG is not supported by the DPO7000 Series Oscilloscopes,
as these models provide the appropriate source signal at the Probe
Calibration output on the front-panel BNC connector.
Ordering Information
TEK-DPG
TEK-DPG
Deskew Pulse Generator Signal Source
Includes: User manual (071-2341-xx).
Recommended Accessories
Tektronix Power Deskew/Calibration Fixture - (067-1686-xx).
Service Options
|
Option
|
Description
|
|
R3
|
Repair Service 3 Years
|
|
R5
|
Repair Service 5 Years
|
|
Additional Service Products Available During
Warranty (DW)or Post Warranty (PW)
|
|
TEK-DPG-R1PW
|
Repair service coverage 1-year post warranty
|
|
TEK-DPG-R2PW
|
Repair service coverage 2-year post warranty
|
|
TEK-DPG-R3DW
|
Repair service coverage 3 years (includes product warranty
period). 3-year period starts at time of customer instrument purchase
|
|
TEK-DPG-R5DW
|
Repair service coverage 5 years (includes product warranty
period). 5-year period starts at time of customer instrument purchase
|
|
Product(s) are manufactured in ISO registered facilities.
|
|
Product(s) complies with IEEE Standard 488.1-1987, RS-232-C,
and with Tektronix Standard Codes and Formats.
|
51W-21697-0, 22-APR-2008
|
|
 |
(WebID: 13429)
|