數位螢光示波器
TDS5034B • TDS5054B • TDS5104B
|
CharacteristicsVertical System | | TDS5034B | TDS5054B | TDS5104B |
|---|
| Input Channels | 4 | | Analog Bandwidth (-3 dB) 5 mV/div - 1 V/div | 350 MHz | 500 MHz | 1 GHz | | Calculated Rise Time 5 mV/div (typical) | 1.15 ns | 800 ps | 300 ps | | Hardware Bandwidth Limits | 150 MHz or 20 MHz | | Input Coupling | AC, DC, GND | | Input Impedance, 1 MΩ | ±1% | | Input Impedance, 50 Ω | ±1% | ±2.5% | | Input Sensitivity, 1 MΩ | 1 mV/div to 10 V/div | | Input Sensitivity, 50 Ω | 1 mV/div to 1 V/div | | Vertical Resolution | 8-bits (>11-bits w/ averaging) | | Max Input Voltage, 1 MΩ | 150 V CAT I, ≤ 400 V peak. Derate at 20 dB/decade to 9 VRMS above 200 kHz | | Max Input Voltage, 50 Ω | 5 V RMS with peaks <±30 V | <100 mV/div <1 VRMS ≥100 mV/div <5 VRMS | | DC Gain Accuracy | 1.5% with offset set to 0 V | | Offset Range, 1 MΩ | 1 mV/div - 99.5 mV/div ±1 V 100 mV/div - 1 V/div ±10 V 1.01 V/div - 10 V/div ±100 V | | Offset Range, 50 Ω | 1 mV/div - 99.5 mV/div ±1 V 100 mV/div -1 V/div ±10 V | 1 mV/div - 50 mV/div ±0.5 V 50.5 mV/div - 99.5 mV/div ±0.25 V 100 mV/div - 500 mV/div ±5 V 505 mV/div - 1 V/div ±2.5 V | | Channel-to-channel Isolation for Any Two Channels at Equal Vertical Scale | ≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth |
Timebase System | Timebase Range | 200 ps/div to 1000 s/div | | Timebase Delay Time Range | (s/div × 10) to 1000 s | | Channel-to-channel Deskew Range | ±75 ns | | Timebase Accuracy | 15 ppm | | Delta Time Measurement Accuracy | (0.06/sample rate + 15 ppm × |Reading|)RMS | | Trigger Jitter (RMS) | 8 psRMS (typical) | | Long Term Sample Rate and Delay Time Accuracy | ±15 ppm over any ≥1 ms interval |
Acquisition System | | TDS5034B | TDS5054B / TDS5104B |
|---|
| Real-time Sample Rates | | | | 1 Channel (max) | 5 GS/s | | 2 Channels (max) | 2.5 GS/s | | 3-4 Channels (max) | 1.25 GS/s | | Equivalent Time Sample Rate (max) | 250 GS/s | | Maximum Record Length per Channel with Standard Memory | 8M/4M/2M-- | 16M/8M/4M | | With Opt. 3M | 16M/8M/4M | NA |
Maximum Duration at Highest Real-time Resolution (1 ch) | | TDS5034B | TDS5054B / TDS5104B |
|---|
| Time Resolution (single shot) | 200 ps (5 GS/s) | | Max Duration with Standard Memory | 1.6 ms | 3.2 ms | | Max Duration with Opt. 3M | 3.2 ms | NA |
Acquisition Modes | FastAcq Acquisition | FastAcq optimizes the instrument for analysis of dynamic signals and capture of infrequent events. Maximum FastAcq waveform capture rate is 100,000 wfms/s. | | Sample | Acquire sampled values | | Peak Detect | Captures narrow glitches (<1 ns) at all real-time sampling rates | | Averaging | From 2 to 10,000 waveforms included in average | | Envelope | From 2 to 2 × 109 waveforms included in min-max envelope | | Hi-Res | Real-time boxcar averaging reduces random noise and increases resolution | | Waveform Database | Accumulates a waveform database that provides a three dimensional array of amplitude, time, and counts | | FastFrame™ Acquisition | Acquisition memory divided into segments; maximum trigger rate >100,000 waveforms per second |
Trigger System | Sensitivity | | Internal DC Coupled | 0.35 div DC to 50 MHz increasing to 1 div at rated bandwidth | | External (auxiliary input) | 400 mV from DC to 50 MHz increasing to 750 mV at 100 MHz | | Main Trigger Modes | Auto, Normal, and Single | | Trigger Sequences | Main, Delayed by time, Delayed by events. All sequences can include separate horizontal delay after the trigger event to position the acquisition window in time. | | Standard Trigger Types | Edge, Glitch, Runt, Window, Width, Transition Time, Timeout, Pattern, Video, State, Setup/Hold | | A Event and Delayed B Event Trigger Types | | | A Event | All above types | | Delayed B Event | Edge | | Communications-related Triggers (requires Option SM) | Support for AMI, HDB3, BnZS, CMI, MLT3, and NRZ encoded communications signals. Select among isolated positive or negative one, zero pulse form or eye patterns as applicable to standard | | Trigger Level Range | | Any Channel | ±10 divisions from center of screen | | External (auxiliary in) | ±8 V | | Line | Fixed at 0 V | | Trigger Coupling | DC, AC (attenuate <60 Hz), HF reject (attenuate >30 kHz) LF reject (attenuates <80 kHz) Noise reject (reduce sensitivity) | | Trigger Holdoff Range | 1.5 μs to 12 s maximum |
Trigger Modes| Edge | Positive or negative slope on any channel or front panel auxiliary input. Coupling includes DC, AC, noise reject, HF reject, and LF reject. | | Video | Trigger on NTSC, PAL, SECAM, analog HDTV, and non-standard video formats | | Glitch | Trigger on or reject glitches of positive, negative or either polarity. Minimum glitch width is 1.0 ns with 200 ps resolution. | | Width | Trigger on width of positive or negative pulse either within or out of selectable time limits ranging from 1 ns to 1 s with 200 ps resolution | | Runt | Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again. Event can be time or logic qualified (logic on 4 channel models only). | | Window | Trigger on an event that enters or exits a window defined by two user-adjustable thresholds. Event can be time or logic qualified (logic on 4 channel models only). | | Timeout | Trigger on an event which remains high, low or either, for a specified time period, selectable from 1 ns to 1 s with 200 ps resolution. | | Transition | Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative or either | | Setup/Hold | Trigger on violations of both setup time and hold time between clock and data present on any two input channels | | Pattern | Trigger when pattern goes false or stays true for specified period of time. Pattern (AND, OR, NAND, NOR) specified for four input channels defined as High, Low or Don't Care. | | State | Any logical pattern of channels (1, 2, 3) clocked by edge on channel 4. Trigger on rising or falling clock edge | | Comm (requires option SM) | Support for AMI, HDB3, B3ZS, B6ZS, B8ZS, CMI, NRZ and MLT3 encoded communication signals. Select among isolated positive or negative one, zero pulse form or eye patterns as applicable to standard. | | Trigger Delay by Time | 16 ns to 250 seconds | | Trigger Delay by Events | 1 to 10,000,000 Events |
Waveform Measurements| Automatic Measurements | 53 of which 8 can be displayed on screen at any one time | | Amplitude Related | Amplitude, High, Low, Maximum, Minimum, Peak to Peak, Mean, Cycle Mean, RMS, Cycle RMS, Positive Overshoot, Negative Overshoot | | Time Related | Rise Time, Fall Time, Positive Width, Negative Width, Positive Duty Cycle, Negative Duty Cycle, Period, Frequency, Delay | | Combination | Area, Cycle Area, Phase, Burst Width | | Histogram Related | Waveform count, Hits in box, Peak hits, Median, Maximum, Minimum, Peak to Peak, Mean (μ), Standard Deviation (σ), μ ± 1σ, μ ± 2σ, μ ± 3σ | | Communications Related | Extinction Ratio (abs, %, dB), Eye Height, Eye Width, Eye Top, Eye Base, Crossing %, Jitter (Pk-Pk, RMS, 6σ), Noise (Pk-Pk, RMS), Signal/Noise Ratio, Cycle Distortion, Q-Factor | | Measurement Statistics | Mean, Min, Max, Standard Deviation, Population | | Reference Levels | User definable for each of the eight measurements | | Histograms | Vertical or horizontal with linear or log scaling | | Gating | Isolate the specific occurrence within an acquisition to take measurements on | | Cursors | Horizontal Bars, Vertical Bars, Waveform, and Screen |
Waveform Processing/Math| Arithmetic | Add, subtract, multiply, and divide waveforms | | Algebraic Expressions | Define extensive algebraic expressions including waveforms, scalars, user adjustable variables, and results of parametric measurements e.g. (Integral (Ch1-Mean(Ch1))*1.414*VAR1) | | Math Functions | Average, Invert, Integrate, Differentiate, Square Root, Exponential, Log 10, Log e, Abs, Ceiling, Floor, Min, Max, Sin, Cos, Tan, ASin, ACos, ATan, Sinh, Cosh, Tanh | | Frequency Domain Functions | Spectral magnitude and phase, real and imaginary spectra | | Vertical Units | Magnitude: Linear, dB, dBm. Phase: degrees, radians, group delay | | Window Functions | Rectangular, Hamming, Hanning, Kaiser-Bessel, Blackman-Harris, Gaussian, Flattop2, Tek Exponential | | Limit Testing | Compare live waveforms against a known “golden” reference waveform with user defined vertical and horizontal tolerances |
Display Characteristics| Display Type | 10.4 in. Liquid crystal active-matrix color display | | Touch Screen | Standard on TDS5054B and TDS5104B. Optional Touch Screen on TDS5034B | | Display Resolution | 640 horizontal x 480 vertical pixels | | Waveform Styles | Vectors, Dots, Intensified Samples, Variable Persistence, Infinite Persistence | | Display Format | YT, XY, XYZ | | Color Palettes | Individual color palettes for Record View and FastAcq/WfmDB modes include Normal, Green, Gray, Temperature, Spectral, and User Defined |
Computer System and Peripherals| Operating System | Windows 2000. | | CPU | Intel Celeron Processor, 2.0 GHz | | PC System Memory | 512 MB | | Internal Hard Disk Drive | ≥80 GB capacity | | CD-RW Drive | Side panel CD-RW drive, ≥24X read and write speed | | Mouse | Optical wheel mouse, USB interface |
OpenChoice Features| TekVISA | Application Programmers Interface (API) for Windows developers. Documentation includes descriptions and samples of programming test and measurement applications on the unit in Visual BASIC, C and C++. | | TekVISA Control (TVC) | Active controls to make access to TekVISA easy for integration into Microsoft Windows applications | | VXI-11 Server | An Application Programmers Interface (API) for LAN connectivity from non-Windows environments | | Plug-and-play Drivers | Provides support to run National Instrument's LabVIEW and LabWindows on an external PC connected to a TDS5000B or on the oscilloscope itself. Instrument drivers are version specific and might not support the version of your software development tools. | | IVI Drivers | Provides support for new and existing program environments utilizing the IVI instrumentation standard, such as LabVIEW, LabWindows/CVI, MATLAB, Visual BASIC, and C/C++. Instrument drivers are version specific and might not support the version of your software development tools. | | Excel and Word Toolbars | Provides direct access to screen images, waveform data, and measurements on the oscilloscope from a toolbar in Excel and/or Word | | Report Generator | Enables the ability to design and create customized report templates that extract the oscilloscope's waveforms, settings, measurements, and other on-screen information with a click of the mouse |
Input/Output Ports| Auxiliary Input | Front panel BNC connector. Trigger level range is adjustable from +8 V to -8 V. The maximum input voltage is ±20 V (DC + peak AC) and input resistance is ≥1.5 kΩ. | | Probe Compensator Output | Front panel pins. Amplitude 1 V ± 1% into a ≥10 kΩ load, frequency 1 kHz ±5% | | Analog Signal Output | Rear-panel BNC connector, provides a buffered version of the signal that is attached to the Channel 3 input (4 channel models only). Amplitude: 50 mV/div ±20% into a 1 MΩ load, 25 mV/div ±20% into a 50 Ω load. Bandwidth (typical): 100 MHz into a 50 Ω load. | | Auxiliary Output Levels | Rear-panel BNC connector, provides a TTL-compatible, negative polarity pulse when the oscilloscope triggers | | External Reference In | Rear-panel BNC connector. 9.8 MHz to 10.2 MHz | | Parallel Port | IEEE 1284, DB-25 connector | | Audio Ports | Miniature phone jacks for stereo microphone input and stereo line output | | USB Port | Four USB 2.0 ports allows connection or disconnection of USB keyboard and/or mouse while oscilloscope power is on | | Keyboard Port | PS-2 compatible | | Mouse Port | PS-2 compatible | | LAN Port | RJ-45 connector, supports 10Base-T and 100Base-T | | Serial Port | DB-9 COM1 port | | Video Port | DB-15 female connector; connect a second monitor to use dual-monitor display mode. Supports basic requirements of PC99 specification and display resolutions up to 1,920 x 1,440 | | GPIB Port | IEEE 488.2 standard, can be configured for talk/listen or controller mode | | Oscilloscope VGA Video Port | DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector |
Power Source | Power | 100 to 240 VRMS ±10%, 47 to 63 Hz; CAT II, or 115 VRMS ±10%, 360 to 440 Hz | | Power Consumption | <220 W |
Physical Characteristics | Configuration | Benchtop | Rackmount | Benchtop | Rackmount |
|---|
| Dimensions | mm | in. | | Height | 361 | 267 | 11.2*1 | 10.5 | | Width | 447 | 483 | 17.6 | 19 | | Depth | 288 | 231*2 | 11.35 | 9.1*2 | | Weight | kg | lbs. | | Net | 11.23 | 13.49 | 24.75 | 29.75 | | Shipping | 25.63 | — | 56.5 | — | | Cooling | | | Cooling clearance | 76 mm required on left side | 3 inches required on left side |
*1 Does not include accessory pouch. *2 From rack mounting rear to back of instrument. Environmental| Temperature | | | Operating | +5 °C to +45 °C | | Nonoperating | -20 °C to +60 °C without diskette in floppy drive | | Humidity | | | Operating | 20% to 80% relative humidity with a maximum wet bulb temperature of +29 °C at or below +45 °C, noncondensing. Upper limit de-rates to 30% relative humidity at +45 °C. | | Nonoperating | Without diskette in floppy disk drive. 5% to 90% relative humidity with a maximum wet bulb temperature of +29 °C at or below +60 °C, noncondensing. Upper limit derates to 20% relative humidity at +60 °C. | | Altitude | | | Operating | 10,000 ft. (3,048 m) | | Nonoperating | 40,000 ft. (12,190 m) | | Random Vibration | | | Operating | 0.1 GRMS from 5 to 500 Hz, 10 minutes each axis, 3-axes, 30 minutes total | | Nonoperating | 2.0 GRMS from 5 to 500 Hz, 10 minutes each axis, 3-axes, 30 minutes total | | Regulatory Certifications | | | Electromagnetic Compatibility | 89/336/EEC | | Safety | UL61010, CSA-22.2 No. 1010.1, EN61010-1, IEC61010-1 |
Ordering InformationTDS5034B350 MHz, 5 GS/s, 4 channel digital phosphor oscilloscope. TDS5054B500 MHz, 5 GS/s, 4 channel digital phosphor oscilloscope TDS5104B1 GHz, 5 GS/s, 4 channel digital phosphor oscilloscope Includes: (1) P5050 500 MHz, 10x passive probe per channel, Accessory Pouch (016–1935–00), Front Cover (200-4651-00), Mouse (119-6936-00), Quick Start User Manual, TDS5000B Product Software CD-ROM, TDS5000B Operating System Restoration CD-ROM, GPIB Programmer's Reference, Optional Applications Software CD-ROM, Getting Started with OpenChoice® book (020-2513-00), Performance Verification Procedure PDF file, Calibration Certificate Documenting NIST Traceability, Z540-1 Compliance, and ISO9001 Registration, Power Cord.Note: Please specify power plug and manual version when ordering. Recommended Probes| P5050 | 500 MHz, 10x passive probe | | P6243 | 1.0 GHz active probe | | P6245 | 1.5 GHz active probe | | P6246 | 400 MHz differential probe | | P6247 | 1.0 GHz differential probe | | P6248 | 1.7 GHz differential probe | | P6250 | DC to 500 MHz, 42 V, Differential Probe (requires TCA-BNC adapter) | | P6251 | DC to 1 GHz, 42 V, Differential Probe (requires TCA-BNC adapter) |
Recommended Accessories| Service Manual | Order 071-1362-00 | | Transit Case | Order 016-1937-00 | | Probe Calibration, Compensation, and Deskew Adapter | Order 067-0405-xx | | Power Deskew Fixture | Order 067-1478-00 | | Mini Keyboard | Order 118-9402-00 | | USB Test Fixture | Order TDSUSBF | | Ethernet Test Fixture | Order through Crescent Heart Software (www.c-h-s.com) |
Instrument Options (Available on all models unless indicated otherwise)| Opt. 18 | Touch-screen interface for TDS5034B | | Opt. 3M | Increase record length to 16M samples max (1ch.) for TDS5034B | | Opt. 1K | Oscilloscope cart | | Opt. 1R | Rackmount kit | | Opt. SM | Communication mask testing | | Opt. VNM | CAN bus decode requires ATM1 trigger module | | Opt. PS1 | Power Bundle that includes TCP202 DC Coupled Current Probe, P5205 High Voltage Differential Probe, TDSPWR3 Power Measurement Software, and Power Deskew Fixture | | Opt. CP2*1 | TDSCPM2 - ANSI/ITU telecom pulse compliance testing software | | Opt. ET3 | TDSET3 - Ethernet compliance test software | | Opt. J2*2 | TDSDDM2 - Disk drive measurements software | | Opt. JA3 | TDSJIT3 v2.0 Advanced - Jitter and timing analysis software | | Opt. JE3 | TDSJIT3 v2.0 Essentials - Jitter and timing analysis software | | Opt. PW3 | TDSPWR3 - Power measurements software | | Opt. USB | USB 2.0 Compliance test software only |
*1 Requires option SM. *2 Not available on TDS5034B. Power Plug Options| Opt. A0 | North America | | Opt. A1 | Universal Euro | | Opt. A2 | United Kingdom | | Opt. A3 | Australia | | Opt. A4 | 240V North America | | Opt. A5 | Switzerland | | Opt. A6 | Japan | | Opt. A10 | China | | Opt. A99 | No power cord or AC adapter |
Manual Options| Opt. L0 | English Manual | | Opt. L1 | French Manual | | Opt. L3 | German Manual | | Opt. L5 | Japanese Manual | | Opt. L7 | Simple Chinese Manual | | Opt. L8 | Traditional Chinese Manual | | Opt. L9 | Korean Manual | | Opt. L10 | Russian Manual |
Service Options| Opt. C3 | Calibration Service 3 Years | | Opt. C5 | Calibration Service 5 Years | | Opt. D1 | Calibration Data Report | | Opt. D3 | Calibration Data Report 3 Years (with Option C3) | | Opt. D5 | Calibration Data Report 5 Years (with Option C5) | | Opt. R3 | Repair Service 3 Years (including warranty) | | Opt. R5 | Repair Service 5 Years (including warranty) |
Instrument UpgradesUpgrades equivalent to original options can be ordered to extend instrument performance after initial purchase. Users can install upgrades without opening the instrument case or requiring on-site service (except for touch-screen upgrade for TDS5034B). To upgrade, order a TDS5BUP with one or more of the following options: 18, 1K, M03, 1R, SM, CP2, ET3, J2, JE3, JA3, PW3, USB, VNM. Factory installation of selected options is available by ordering option IF on your TDS5BUP upgrade order.
| Product(s) are manufactured in ISO registered facilities. |
55W-14869-13
|
|
 |
(網站 ID: 13456)
|