Electrical Clock Recovery Modules
80A05/80A07 Modules for DSA8200 Series Oscilloscopes
|
Features & Benefits
- Electrical Clock Recovery for:
- Enumerated Bit Rates between 50 Mb/s and 12.6 Gb/s
- Continuously Variable Rates from 100 Mb/s to 12.5 Gb/s Enables
Support for Current, Emerging, and Future Bit Rate Needs
- High Jitter Tolerance, Slow Slew Rate Tolerance for Recovering
Clocks from Stressed, Degraded, or Spread-spectrum Clocked Signals
- Excellent Time Base Stability and Ultra-low Residual Jitter
(as low as 250 fsRMS) for Best Measurement System Fidelity
- Unequaled PLL Bandwidth and Roll-off Shape Control (Tunable
loop bandwidth of 100 kHz to 12 MHz) Enables Control Required by Many
Compliance Standard Tests
- Clean 50 Ω Path for the Best Signal
Acquisition Fidelity
Applications
- High-speed Serial Data Link
and Device Characterization for Computer, Communications, and Consumer
Applications
- Compliance Testing of Electrical Signaling
- High-speed Optical Communications Testing
- Jitter, Noise,
BER, and Signal Impairment Analysis
- Testing of Frequency
Agile or Multirate Devices
The 80A05 and 80A07 Electrical Clock Recovery Modules enable clock
recovery for electrical signals. Additionally, the 80A05 provides
internal routing of the recovered clock for triggering of the 8200*1 Series oscilloscopes.
The 80A05 recovers clocks
from serial data streams for all of the most common electrical standards
in the 50 Mb/s to 3.188 Gb/s range (continuous coverage) plus the
(fixed) rate of 4.25 Gb/s. Option 10G adds support for user-selectable
rates in the following ranges:
- 3.267 to 4.250 Gb/s;
- 4.900 to 6.375 Gb/s; and
- 9.800 to 12.60 Gb/s range
The 80A05 provides high level of integration and the best
sensitivity available, and is the optimal solution for testing of
optical transmitter components and electrical components of optical
systems. The 80A05 PLL loop bandwidth selection is automatic with
no interaction, leading to easy setup and operation.
The 80A07
recovers clocks from serial data streams for all of the most common
electrical standards in the continuous 100 Mb/s to 12.5 Gb/s range.
Auto-locking capability is selectable from the user interface or programmatic
interface, so the design and test engineers can search and lock onto
signals of unknown data rates. The 80A07 offers complete configurability
and state-of-the-art specifications and is the preferred solution
for most serial data standards due to excellent stability, superior
Jitter and slew rate tolerance for recovering clocks from stressed
or degraded signals, and unequaled PLL bandwidth and roll-off shape
control for either Golden PLL compliance testing or custom PLL response.
The 80A07 also locks on spread-spectrum signals.
In addition
to standard rate support, users can specify custom bit rates on either
of the two modules in order to test devices, modules, and systems
running at emerging or nonstandard rates. The wide clock recovery
ranges support for user-specified bit rates provide complete clock
recovery solutions for testing a wide range of computer, communications,
and consumer electronics signaling rates and standards.
The
80C12 optical module with either the 80A05 or 80A07 clock recovery
modules provides a complete solution for optical rates between 155 Mb/s
and 12.6 Gb/s.
*1 Also compatible with CSA/TDS8200,
CSA/TDS8000B, and CSA/TDS8000 sampling oscilloscopes.
Both
modules accept either single-ended or differential signals at their
inputs. With either single-ended or differential signals, the attenuated
but otherwise unmodified input signal is available on output connectors
on the front panel of the modules. The signal path to these front-panel
outputs has been carefully designed to preserve signal fidelity well
beyond the frequency corresponding to the maximum bit rate addressed
by the clock recovery circuit. The front-panel output signals can
therefore be connected to a high-frequency sampling module (such as
80E07) and be acquired for analysis while preserving high-frequency
features of the signal.
Tektronix clock recovery solutions
combine simplicity of use with excellent flexibility; the full rate
recovered clock or its sub-rate is available on the modules’ front
panel to clock or trigger other equipment.
Characteristics
Supported Specifications
|
Specification
|
80A05
|
80A07
|
|
Standard
|
Option 10G
|
|
OC3/STM1
|
155.52 Mb/s
|
▀
|
▀
|
♦*2
|
|
OC12/STM4
|
622.08 Mb/s
|
▀
|
▀
|
▀
|
|
Fibre Channel
|
1.063 Gb/s
|
▀
|
▀
|
▀
|
|
Gigabit Ethernet
|
1.25 Gb/s
|
▀
|
▀
|
▀
|
|
SATA Gen I
|
1.50 Gb/s
|
♦*3
|
♦*3
|
▀
|
|
2 GB Fibre Channel
|
2.125 Gb/s
|
▀
|
▀
|
▀
|
|
OC48/STM16
|
2.488 Gb/s
|
▀
|
▀
|
▀
|
|
2 GB Ethernet
|
2.50 Gb/s
|
▀
|
▀
|
▀
|
|
PCI Express I
|
2.50 Gb/s
|
♦*3
|
♦*3
|
▀
|
|
Infiniband®
|
2.50 Gb/s
|
▀
|
▀
|
♦*2
|
|
2.5G G.709 FEC
|
2.666 Gb/s
|
▀
|
▀
|
♦*2
|
|
SATA Gen II
|
3.0 Gb/s
|
♦*3
|
♦*3
|
▀
|
|
XAUI, 10GBase-X
|
3.125 Gb/s
|
▀
|
▀
|
♦*2
|
|
10 GB FibreChannel x4
|
3.188 Gb/s
|
▀
|
▀
|
♦*2
|
|
4 GB FibreChannel
|
4.25 Gb/s
|
▀
|
▀
|
▀
|
|
FB-DIMM1
|
3.2, 4.0, 4.8 Gb/s
|
|
♦*2, 3
|
▀
|
|
PCI Express II
|
5.0 Gb/s
|
|
♦*2, 3
|
▀
|
|
FB-DIMM2
|
4.8, 6.4, 8.0, 9.6 Gb/s
|
|
♦*2, 3
|
▀
|
|
OIF CEI
|
6+ Gb/s
|
|
♦*2
|
▀
|
|
2x XAUI
|
6.25 Gb/s
|
|
▀
|
♦*2
|
|
8 GB Fibre Channel
|
8.50 Gb/s
|
|
|
▀
|
|
OC192/STM64
|
9.953 Gb/s
|
|
▀
|
▀
|
|
XFP/XFI
|
9.95-11.2
|
|
♦*2
|
▀
|
|
10 G Base-W
|
9.953 Gb/s
|
|
▀
|
|
|
10 G Base-R
|
10.31 Gb/s
|
|
▀
|
▀
|
|
10 GB Fibre Channel
|
10.51 Gb/s
|
|
▀
|
♦*2
|
|
G.975 FEC
|
10.66 Gb/s
|
|
▀
|
♦*2
|
|
G.709 FEC
|
10.71 Gb/s
|
|
▀
|
♦*2
|
|
OIF CEI
|
11+ Gb/s
|
|
|
▀
|
|
10 GbE w/ FEC
|
11.10 Gb/s
|
|
▀
|
♦*2
|
|
Super FEC
|
12.50 Gb/s
|
|
▀
|
♦*2
|
|
|
|
Additional enumerated standard rates are supported
with 8200*1 Series firmware releases higher than 2.4.x
|
|
|
Rates Supported: ▀ = Filter, ♦ = Optical Clock
Recovery
|
*2 The standard is not enumerated, but
is supported as a custom rate.
*3 No spread-spectrum
clocking support.
Clock Recovery Ranges for custom (user
specified) rates (in addition to previous enumerated
lists)
|
80A05
|
80A07
|
|
Standard
|
Option 10G
|
|
50 Mb/s to 3.188 Gb/s
4.25 Gb/s
|
50 Mb/s to 3.188 Gb/s
3.267 to 4.25 Gb/s
4.900
to 6.375 Gb/s
9.800 to 12.60 Gb/s
|
100 Mb/s to 12.5 Gb/s continuous
|
Sensitivity: (clock recovery will lock, differential
data is given for each input)
|
|
80A05
|
80A07
|
|
Standard
|
Option 10G
|
|
Lowest Supported Rate to 2.70 Gb/s
|
Differential ≤8 mVp-p
Single
ended 10 mVp-p
|
Differential 15 mV (typical)
Single ended
30 mV (typical)
|
|
2.70 to 11.19 Gb/s
|
|
Differential ≤12 mVp-p
Single ended 15 mVp-p
|
|
11.19 to 12.60 Gb/s
|
|
Differential ≤15 mVp-p
Single ended 20 mVp-p
|
Sensitivity: (clock recovery will lock, differential
data is given for each input) (Continued)
|
Characteristic
|
80A05
|
80A07
|
|
Standard
|
Option 10G
|
|
I/O
|
|
I/O Connectors
|
SMA
|
APC 3.5
|
|
I/O Impedance
|
50 Ω
|
50 Ω
|
|
Input
+Data In and -Data In; complementary signals
|
|
Input Voltage:
|
|
Absolute Maximum Nondestructive
|
2.5 Vp-p (each input)
|
±5 V maximum
|
|
Absolute Maximum Operational
|
2.0 Vp-p (either input)
1.0 Vp-p (differential, each input)
|
|
|
Maximum Input Signal Skew (+Data In to -Data In under which the unit will still meet its sensitivity specification)
|
50 Mb/s to 2.70 Gb/s:
20% of unit interval
2.70 to 12.6 Gb/s:
20 ps
|
100 Mb/s to 2.70 Gb/s: 20% of unit interval
2.70 to 12.6 Gb/s: 20 ps
|
|
Measured Edge Density
|
N/A
|
±1%
|
|
Measured Phase Deviation
|
N/A
|
Displayed as %RMS and %p-p, 10-90%p-p available range
|
|
Coupling
|
|
|
|
Recovered Clock Output (to Trigger Clock output)
|
AC
|
AC
|
|
Data Input to
Data Out
|
DC
|
AC
|
|
Output
+Data Out and -Data Out; complementary signals
|
|
Attenuation
+Data In to
+Data
Out
-Data In to
-Data Out
|
6.6 dB ±0.6 dB
(ratio of +Data Out / +Data In )
|
2 dB (min)
2.6 dB (typical)
3 dB (max)
|
|
Bandwidth (-3 dB)
+Data In to
+Data Out
-Data In to
-Data
Out
|
≥20 GHz
|
≥20 GHz
|
Recovered TRIGGER CLOCK (80A05) / CLOCK (80A07)
Output
|
Characteristic
|
80A05
|
80A07
|
|
Standard
|
Option 10G
|
|
Output Frequency
|
|
Input Bit Rate <2.70 Gb/s
|
Input bit rate
|
Full rate divided by 1, 2, 4, 5, 6, 7, 8, 9, 10, 12, 14, 16,
18, 20, 24, 25, 28, 30, 32, 35, 36, 40, 42, 45, 48, 49, 50, 54, 56,
60, 63, 64, 70, 72, 80, 81, 90, 100, 108, 112, 120, 126, 128, 140,
144, 160, 162, 168, 180, 192, 196, 200, 216, 224, 240, 252, 256, 280,
288, 320, 324, 336, 360, 384, 392, 432, 448, 504, 512, 576, 648. Odd
division ratios have non-50% duty cycles.
|
|
Input Bit Rate ≥2.70 Gb/s
|
Input bit rate / 16
|
|
|
Loop Bandwidth
|
BW of (bit rate / 1666) till 3 Gb/s
4 MHz
nominal at other ranges and BW <4 MHz at 9.953 and 10.3125 Gb/s
|
100 kHz – 12 MHz variable
(see CRU Loop Bandwidth graph
on next page for calibrated loop bandwidth region)
|
|
Loop Bandwidth Accuracy
|
Nominal
|
±5%
|
|
Locking Range
|
1000 ppm nominal
|
50 MHz, adjustable to 10-500 MHz
|
|
Peaking
|
N/A
|
0-6 dB from 500 kHz to 12 MHz
0 dB from 100 to 500 kHz
|
|
Peaking Accuracy
|
N/A
|
Greater of ±10% of settings, or 0.2 dB
|
|
Jitter
|
|
155 Mb/s to 2.70 Gb/s
|
0.5% of unit interval (RMS) typical
|
70 fsRMS (typical), 250 fsRMS (max)
(measured at 800 mVp-p input amplitude, 10 Gb/s, 1010
pattern, 2 MHz loop bandwidth setting and 0.5 dB peaking)
|
|
2.70 to 6.38 Gb/s
|
1.27 psRMS typical
|
|
|
9.80 to 12.6 Gb/s
|
0.6psRMS typical
|
|
|
Return loss
|
DC to 10 GHz: 15 dB
10 GHz to 20 GHz: >10 dB
|
15 dB
|
|
Rise/Fall Times
|
TRIGGER CLOCK output: <300 ps typical (10-90%)
|
10 G CLOCK output: <30 ps typical (10-90%)
|
25 ps (typical), 30 ps (max) (20/80%)
|
|
Amplitude
|
>400 mV (typical)
|
250 mV (min), 1.5 V (max), 900 mVp-p (typical),
accuracy greater of 10% or 30 mV
|
|
Output Frequency Deviation Tracking Range (Tracking 30 to
33 kHz Triangle Modulated SSC)
|
N/A
|
+500/-5500 ppm
(+0.05/-0.55%)
|
Recovered 10G CLOCK OUT (Option 10G only)
|
Characteristic
|
80A05
|
80A07
|
|
Standard
|
Option 10G
|
|
Jitter
|
|
2.70 to 3.14 Gb/s
|
N/A
|
4x input bit rate; 2.5 psRMS
|
N/A
|
|
3.27 to 4.25 Gb/s
|
3x input bit rate; 2.5 psRMS
|
|
4.90 to 6.38 Gb/s
|
2x input bit rate; 2.5 psRMS
|
|
9.80 to 12.6 Gb/s
|
1x input bit rate; <2.0 psRMS
|
|
Amplitude
|
|
>500 mV (typical)
|
|
|
Trigger Output (80A07 Rear panel) (External lock
status indication)
|
|
Interface Type
|
N/A
|
SMA, 50 Ω DC coupled
|
|
Latency
|
N/A
|
300 ms
|
|
Trigger Input (80A07 Rear panel) (External clock
recovery lock request)
|
|
Interface type
|
N/A
|
SMA, 50 Ω DC coupled to 0 V
|
|
Threshold
|
N/A
|
1.5 V
|
|
Minimum Pulse Width
|
N/A
|
50 ns
|
Physical Characteristics
|
Dimension
|
80A05
|
80A07
|
|
Width
|
6.5 in. / 165 mm
|
15.5 in. / 394 mm
|
|
Height
|
1.0 in. / 25 mm
|
3.75 in. / 95 mm
|
|
Depth
|
12.0 in. / 305 mm
|
13.75 in. / 337 mm
|
|
Weight
|
2.7 lb. / 1.22 kg
|
20 lb. / 9.1 kg
|
80A07 Clock Recovery Loop Bandwidth vs. Data Rate
The 80A07 clock recovery module has a variable loop
bandwidth from 100 kHz to 12 MHz. The calibrated loop bandwidth is
limited as shown on the graph above.
Ordering Information
80A05
Multirate
Electrical Clock Recovery Module
Includes: User Manual.
Note: Also used for 80C12 Optical Sampling Module Clock
Recovery.
Product Options
Opt. 10G - Add bit
rates: 3.267 Gb/s to 4.25 Gb/s.
4.900 Gb/s to 6.376 Gb/s.
9.800 Gb/s to 12.60 Gb/s.
Service Options
Opt. C3 - Calibration Service 3 Years.
Opt. C5
- Calibration Service 5 Years.
Opt. D1 - Calibration
Data Report.
Opt. D3 - Calibration Data Report (with
C3 only).
Opt. D5 - Calibration Data Report (with C5
only).
Opt. R3 - Extend repair warranty to 3 Years.
Opt. R5 - Extend repair warranty to 5 Years.
Firmware support - This module is supported on 8200*1 Series oscilloscopes running Firmware release 2.0.1.5 or later.
80A07
Multirate clock recovery module.
Includes: User manual, Application CD (includes virtual front-panel application
and remote control software developers kit), USB cable, US power cord,
SMA terminations for data outputs, and quick start guide. A replacement
fuse is also included.
Power Requirements
Line
Voltage and Frequency - 100 to 240 VAC 50/60 Hz.
Rear Panels
Environmental Characteristics
Temperature -
Operating: 0 to 40 °C.
Nonoperating:
-18 to 60 °C.
Relative Humidity -
Operating:
20 to 80% at or below 40 °C.
Nonoperating: 5 to 90% at or below
60 °C.
Vibration (5 Hz to 500 Hz) -
Operating:
0.2 gRMS.
Nonoperating: 2.09 gRMS.
Service Options
C3 - Calibration Service
3 Years.
C5 - Calibration Service 5 Years.
R3 - Repair Service 3 Years.
R5 - Repair Service
5 Years.
CA1 - Provides a single calibration event
or coverage for the designated calibration interval, whichever comes
first.
International Power Plug Options
Opt. A0
- North America power.
Opt. A1 - Universal EURO
power.
Opt. A2 - United Kingdom power.
Opt.
A6 - Japan power.
Opt. A10 - China power.
Interconnect Cables
015-0560-xx - (450 mm/18 in.;
1 dB loss at 20 GHz) cable is a high-quality cable recommended for
work to 20 GHz.
Interconnect Cables (3rd party)
Tektronix recommends using quality high-performance interconnect
cables with Tektronix high-bandwidth products in order to minimize
measurement degradation and variations. The W.L. Gore & Associates'
cable assemblies are compatible with the 2.92 mm, 2.4 mm, and 1.85 mm
connector interface. Assemblies can be ordered by contacting Gore
by phone at (800) 356-4622, or on the web at www.gore.com/tektronix.
Firmware Support
The 80A07 is a standalone clock
recovery module and as such has its own front-panel controls. In addition,
it ships with a virtual front-panel control application which can
be installed on 8200*1 Series sampling oscilloscopes running
version 2.0 or later or on a Windows PC running Windows 2000 or Windows
XP. The controlling processor (8200*1 Series oscilloscope
or Windows PC) must be connected to the 80A07 by a USB cable in order
for this virtual front-panel control application to work.
When
running on an 8200*1 Series sampling oscilloscope or Windows
PC, the virtual front-panel control application can be started from
the Windows desktop or start menu. Additionally, if the 8200*1 Series sampling oscilloscope is running firmware version
2.4 or later, this control application can be invoked using the Apps menu within the sampling oscilloscope application.
Remote control of the 80A07 is available with a software development
kit which comes standard with the 80A07.
*1 Also
compatible with CSA/TDS8200, CSA/TDS8000B, and CSA/TDS8000 sampling
oscilloscopes.
|
|
Product(s) are manufactured in ISO registered facilities.
|
|
Product(s) complies with IEEE Standard 488.1-1987, RS-232-C,
and with Tektronix Standard Codes and Formats.
|
85W-17380-6, 09-MAR-2007
|
|
 |
(WebID: 13579)
|