Features & Benefits
- 18 GHz Bandwidth
- Direct
In-circuit TDR Testing
- Reliable and Repeatable Resolution
of Impedance Variations to 0.12″
- Adjustable Probe Pitch from
0.5 mm to 4.2 mm
- Probing of Fully Balanced Differential Signals
Without Ground Contact
- Double-cushioning Design Enables Probe
Tips to Maintain Proper DUT Contact
- Protects Sampling Module
from Static Damage when Used with Two 80A02 Modules
Applications
- Circuit-board Controlled
Impedance Testing (TDR)
- Loss Tangent Measurements for Differential
Transmission Lines
- Failure Analysis of PCB With or Without
Components Mounted
- Package Impedance Testing (TDR)
P80318
The P80318 is an 18 GHz 100 Ω input impedance differential TDR
hand probe. This probe enables high-fidelity impedance measurements
of differential transmission lines. The adjustable probe pitch enables
a wide variety of differential line spacing and impedances. The P80318
probe also includes two precision SMA cables with parallel control
lines that provide the 80A02 the control for EOS/ESD protection.
P80318X
The P80318X is designed for high-volume PCB
impedance test applications. These applications require maximum manufacturing
up-time. The P80318X includes everything in a P80318 plus a second
probe body for use if the first probe body needs to be serviced due
to normal wear.
The P80318 probe can be used as a standalone
high-resolution probe, but it is specially designed to be used with
two 80A02 EOS/ESD Protection Modules. To maximize testing up-time
and minimize repair costs, protecting the sampling module from static
damage is a requirement. Damage to the sampling module is typically
due to electro-overstress (EOS) and electrostatic discharge (ESD)
from large static charges stored in the DUT. Use of two 80A02 modules
ensures proper probe grounding prior to switching signal to the sampling
module.
Characteristics
Nominal Length with Cable
- 1 meter
Attenuation - 1x
Probe-only
Bandwidth - >20 GHz
Probe Pitch - 0.5 to 4.2 mm
(signal tip to signal tip)
Connector Type - SMA
Measured Reflected TDR Fall Time - <39 ps
Impedance - 50 Ω odd mode, 100 Ω differential
Max
Vin - 5.0 V
Probe-only Propagation Delay
- 1.0 ns typical
TDR Control Activation Force - 150 to 200 grams each probe half
Note: Numeric values shown
are typical
Ordering Information
P80318
18 GHz
100 Ω Differential TDR Hand Probe.
Includes: 18 GHz
bandwidth 50 Ω odd mode input impedance (100 Ω differential impedance),
differential passive handheld probe with one set of two precision
SMA cables, soft storage case, user manual, and three replaceable
ground shorting straps.
P80318X
18 GHz 100 Ω Differential
TDR Hand Probe for Manufacturing.
Includes: Two 18 GHz
bandwidth 50 Ω odd mode input impedance (100 Ω differential impedance),
differential passive handheld probes with one set of two precision
SMA cables, soft storage case, user manual, and three replaceable
ground shorting straps. Second probe body is provided as a backup
to the first to maximize manufacturing up-time if the first probe
body needs to be serviced due to wear.
Recommended Accessories
P8018 - 18 GHz Single-ended 50 Ω TDR Probe. To compliment
TDR probing for single-ended applications.
80A02 - ESD/EOS TDR protection module.
External probe power cable
- To reduce module acquisition slot consumption the 80A02 modules
can run off this probe power cable (p/n 174-3896-xx).
Ground Spring Kit - Set of two ground springs with a
spare set screw for P80318 probe (p/n 020-2670-xx).